10. Measuring using Measuring-Menus
Measuring point selection:
The first measuring channel is always the selected measuring point.
This can be selected directly by means of:
or
...
To change the other channels, the measuring point
PROG
and
must be selected as function, by means of keys:
or
...
The channel can now be changed by means of:
<M
>
,
<M
>
...
The channel selection is terminated with key:
<ESC>
To
set the display range
for the bar chart the functions
"Analog start"
and
"Analog end"
in the
"Special functions" menu should be used;
(see
11.4.4).Having selected these functions they can be entered by pressing
PROG
and
or directly on the appropriate axis; (see 9.5).
10.5.2 Differential Measurement
If there are two sensors with the same dimensions and same decimal point po-
sition connected at measuring points M0 and M1, the difference M1-M0 ap-
pears automatically at the device-internal measuring point M5 (s. 8.2). If the
differential channel is not required, it must be explicitly deleted (s. 11.3.9). If
further differential channels are needed, these can be created using the appro-
priated reference channels (s. 11.4.6).
10.5.3 Menu List of Measuring Points
The best overview of the meas. system incl.
all meas. values, time of day, date and cycle
is obtained with the menu
Meas.point list
.
From here, it is also possible to get to the
Sensor programming
of all measuring points.
This menu cannot be individually configured,
it can only be combined with some selected
functions:
At first call a list with max 20 meas. pts is displayed:
MeasptsList: 20 Meas.v.
00: 23.12°C ...
Functions can be allocated to a meas. value with:
or
...
This reduces the max. number of channels to 10.
Select the following function with:
Measured value including
comment:
MeasValList: Comment
00: 23.12°C Temperature
Measured value including
maximum value:
MeasValList: Max value
40 ALMEMO
®
2890-9
C © REC COM l© ©l R01 * ´´´´´´µµµµ¶
±±±±±±±±±±±±±±±±±±±±±±±±±±±±±±±±±±±±
Meas.points list: Comment
Time: 12:34:56 Date: 01.01.04
Cycle-timer:
00:00:30 nS
00: 23.12 °C Temperatur
01: 11.37 mls Velocity
02: 123.4 mV U2.4
10: 53.6 %H r.Humidity
20: 15.2 °C Dew point
30: 11.2 glk a.Humidity
START MANU F PRINT ESC