15. Output modules
70
The following trigger functions
can be programmed as variants :
0: Start / stop a measuring operation
0: Start - Stop
1: Once-only manual measuring point scan
1: Once-only scan
2: Clear all maximum values and min values
2: Clear max and min values
3: Print measured value
3: Print
4: Start / stop a measuring operation on level-controlled basis
4: Start - Stop, Level-con-
trolled
8: Zero-set a measured value
8: Zero-set meas. value
-5: Execute macro 5 (see manual 6.6.5)
-5: Macro 5
-6: Execute macro 6
-6: Macro 6
-7: Execute macro 7
-7: Macro 7
-8: Execute macro 8
-8: Macro 8
-9: Execute macro 9
-9: Macro 9
15.3
Analog outputs
V5 output modules
For the purposes of analog recording of measured
values it is still possible, at sockets A1 and / or A2
(2)
,
to connect V5 output modules with an analog output
controlled by the device, e.g. recording cable
ZA1601-RK (see Manual 5.1.1).
To select the socket
press :
<P>
:
▲
or
▼
The following output modes
can be programmed as variants :
0: Measured value of selected measuring channel :
0: Selected meas channel M00
2: Measured value of a programmed channel :
2: Assigned internally M01
8: Programmed analog output (see below) :
8: Driven externally
The analog value appears below this in digits :
Analog value : 08345
This gives, depending on the analog output, the following
output signals
:
Voltage output
-1.2 ... +2.00 V
0.1 mV / digit
Voltage output
-6.0 ... +10.0 V
0.5 mV / digit
Current output
0.0 ... 20.0 mA
1 µA / digit
In variant 2 ´Assigned internally´, after selecting the Mxx function,
it is possible to program the measuring point to be output :
2: Assigned internally
M
V6 output modules
The new V6 relay trigger analog adapter ZA8006-
RTA3 (see Manual 5.1.3) offers the option at ports
4 to 7 of up to 4 in the output signal separately
configurable external analog outputs.
02