background image

5. Introduction 
 

10 

characteristic,  e.g.  10  thermocouple  types,  NTC  and  PT100  probes,  infrared 

sensors,  and  flow  transducers  (rotating  vanes,  thermoanemometers,  Pitot 

tubes). For humidity sensors additional function channels are available for cal-

culating humidity variables such as dew point, mixture ratio, vapor pressure, and 

enthalpy. Even complex chemical sensors are supported. Measured values from 

other sensors can also be acquired using the voltage, current, and resistance 

ranges  with  individual  scaling  in  the  connector.  Existing  sensors  can  also  be 
used  -  so  long  as  the  appropriate ALMEMO®  connector  is  connected  via  its 

screw terminals. For digital input signals, frequencies, and pulses, adapter con-

nectors  are  available  with  an  integrated  microcontroller.  It  is  thus  possible  to 
connect  virtually  any  sensor  to  any ALMEMO®  measuring  instrument  and  to 

change sensors without the need for any extra settings. 

Function channels 

Maximum,  minimum,  average,  and  differential  values  from  certain  measuring 

points can be programmed as  function channels, also internal channels, and 

can be processed like normal measuring points. 

Units 

The 2-character units display can be adapted for each measuring channel so 

that  both  the  display  and  via  interface  always  indicate  the  correct  units,  e.g. 

when a transmitter is connected. Conversion between °C (Centigrade) and °F 

(Fahrenheit) is performed automatically. 

Measured value designation 

Each sensor is identified by means of a 10-character alphanumeric name. This 

name is entered via the interface and will appear via interface and in the soft-

ware. 

Correction of measured values 

The measured value on each measuring channel can be corrected both in terms 

of zero-point and gain; this means that even sensors usually requiring initial ad-

justment (e.g. expansion, force, pH) can be freely interchanged. Zero-point cor-

rection and, partly at least, gain adjustment can be performed at the touch of a 

button. Sensors with multi-point calibration can also be connected; (see Manual 

Section 6.3.13). 

Scaling 

The corrected measured value on each measuring channel can also be further 

scaled in terms of zero-point and gain - using the base value and factor. The dec-

imal point position can be set by means of the exponent function. 

Limit values and alarm 

Per measuring channel two limit values can be set (1 maximum and 1 minimum). 

In the event of one of these limit values being exceeded relay output modules 

actuate  the  associated  alarm  contacts;  these  can  be  allocated  individually  to 

specific limit values. Hysteresis is set by default to 10 digits but this can be ad-

justed to any number between 0 and 99. The exceeding of a limit value can also 

be used to start or stop measured value recording automatically. 

Summary of Contents for ALMEMO 2490-1A

Page 1: ...____________________________ Operating instructions English Universal measuring instruments ALMEMO 2490 1A 2490 2A V2 9 04 11 2021 www ahlborn com...

Page 2: ...r isol ZA2690 UK 12V ALMEMO connector ZA1000 FSV 5 LCD a Function b Measuring point 2nd meas value c Units for 2nd measured value d Units for 1st measured value e 1st measured value f Operational stat...

Page 3: ...operations 11 5 1 3 Process control 12 6 Initial commissioning 13 7 Power supply 14 7 1 Battery operation and 14 7 2 Mains operation 14 7 3 External DC voltage supply 14 7 4 Sensor supply 14 7 5 Swit...

Page 4: ...surement 25 11 Outputs 26 11 1 Interface 26 11 2 Analog outputs 26 12 Device configuration 27 12 1 Device address and networking 27 12 2 Analog output 27 12 3 Automatic switch OFF 28 12 4 Device locki...

Page 5: ...ous quality tests We provide a warranty lasting two years from delivery date that your device will function trouble free Before you send your device to us please ob serve the advisory notes in Chapter...

Page 6: ...ol Disposal of any such item as unsorted domestic waste is strictly forbidden Please dispose of all packaging materials according to the applicable national waste management regulations Please dispose...

Page 7: ...unction with unsuitable peripheral equipment Use of the device for any purpose other than that for which it is intended Damage caused by electrostatic discharge or lightning DANGER Risk of fatal injur...

Page 8: ...eir own integrated power supply are not electrically iso lated from one another 4 2 Handling batteries rechargeable batteries cor rectly When inserting batteries rechargeable batteries ensure that the...

Page 9: ...Ch 7 The operating instructions you are now reading cover only those features and controls that are specific to this device Many sections therefore also refer to the more detailed description in the M...

Page 10: ...rect units e g when a transmitter is connected Conversion between C Centigrade and F Fahrenheit is performed automatically Measured value designation Each sensor is identified by means of a 10 charact...

Page 11: ...ach measuring point can be scaled by means of analog start and analog end in such a way that the measuring range thus defined covers the full range of the analog output 2 V 10 V or 20 mA At the analog...

Page 12: ...it data output from particular channels this may be necessary in order to reduce excessive data flow especially while data is being saved Averaging over measuring point scans The measured values from...

Page 13: ...tment by pressing key CLR 6 return to normal measured value by pressing and holding down key CLR see 10 4 7 Differential measurement 2490 2A only plug 2 sensors of same type into sockets M0 and M1 and...

Page 14: ...al DC voltage supply The DC socket 4 can also be used to connect DC voltage 12 V minimum 200 mA This is via an ALMEMO connector ZA1000 FSV If however the power supply has to be electrically isolated f...

Page 15: ...PROM on the sensor connector and the device s calibration and programmed parameters are stored in the EEPROM on the instrument itself both on a fail safe basis Date and time of day are retained intact...

Page 16: ...ary provide up to 4 channels The ad ditional channels can be used in particular for humidity sensors with 4 measur ing variables temperature humidity dew point mixture ratio or for function channels E...

Page 17: ...cally isolated by means of photovoltaic relays and a potential difference of maximum 50 VDC or 60 VAC is permissible between them Sensors combined within one connector and sensors with their own power...

Page 18: ...ain field Display of measuring functions in the function field Measuring point Maximum value Minimum value Saved value Memory capacity Temperature value from double sensors Configuration of device add...

Page 19: ...value flashes Undershoots measuring range Minimum value flashes 9 2 Keypad To operate the device a keypad with 7 keys is provided Function Key To switch ON the device see Section 7 5 ON OFF To switch...

Page 20: ...l operating states possible for the meas ured value are explained in Section 9 1 10 1 1 Selecting a measuring point By pressing key M you can select one after the other all active measuring points and...

Page 21: ...FrEq Pulse ZA 9909 AK2 0 65000 PULS Digital input ZA 9000 EK2 0 0 100 0 Inp Digital interface ZA 9919 AKxx 65000 65000 diGi Snap on head normal 20 FV A915 S120 0 30 20 00 m s S120 Snap on head normal...

Page 22: ...FS C UbGt Measured value of Mb1 any MESS Cold junction temperature any C CJ Number of av values of Mb1 any n t Volume flow m3 h M Mb1 Q any mh FLou Timer any s tinE TC Temperature compensation PC Air...

Page 23: ...n for about one second e g P02 The value most recently saved then appears in the function field preceded by the symbol M To return to the channel display press key M To display all memory data press a...

Page 24: ...d L890 units Pa the zero point must be temporarily adjusted by press ing key CLR i e until switching off even if the channel is locked With the following chemical probes automatic two point adjustment...

Page 25: ...value 7 00 and gain 0 1689 Temperature compensation On conductivity probes and O2 probes with an integrated temperature sensor temperature compensation is performed automatically For pH probes a tempe...

Page 26: ...ing instruments even if these are located far apart see Manual 5 3 The data cables required for this purpose see Manual 5 2 are plugged into socket A1 The baud rate for all data cables is is programme...

Page 27: ...M To delete parameters press CLR select the next digit or entry is completed ON To terminate or cancel configuration MEM 12 1 Device address and networking To communicate with networked devices it is...

Page 28: ...reference channel press key M M To enter a value first press ON and the 1st digit starts flashing Each digit can be changed by pressing keys M or M To use negative values below zero press M To delete...

Page 29: ...rmal pressure 1013 mbar involve certain measuring errors e g Error per 100 mbar Compensation range Rel humidity psychrometer approx 2 500 to 1500 mbar Mixture ratio capacitive approx 10 Vapor pressure...

Page 30: ...ect the sensors again one at a time and check successively If a fault persists for any one connection then check all wiring if nec essary insulate the sensor and eliminate interference by using shield...

Page 31: ...ssed wires Are all network distributors supplied with power Network the devices again one at a time and check successively see above If after performing the above listed checks and remedial steps the...

Page 32: ...32 14 DECLARATION OF CONFORMITY...

Page 33: ...lue 5x 7 segment 15 mm 2x 16 seg ment 9 mm Function 4 x 7 segment 9 mm 9 symbols Operation 7 silicone keys Memory 99 measured values on the RAM Date and time of day Software clock buffered by battery...

Page 34: ...490 2A Accessories Top hat rail mounting OA 2490 HS Rubberized impact protection green ZB 2490 GS1 Mains adapter with ALMEMO connector 12 volts min 1 A ZA 1312 NAx DC adapter cable 10 to 30 V DC 12 V...

Page 35: ...sured value memory 23 Measuring inputs 16 Measuring inputs 33 Measuring operations 20 Measuring ranges 20 memory data 23 minimum value 22 networking 27 OFF 14 ON OFF 19 Operating controls 2 operating...

Page 36: ...Mess und Regelungstechnik GmbH Eichenfeldstra e 1 83607 Holzkirchen Germany internet http www ahlborn com e mail amr ahlborn com Despite great care incorrect information cannot be ruled out Technical...

Reviews: