Ahlborn ALMEMO 2390-3 Operating Instructions Manual Download Page 7

Differential measurement

It is possible, by setting the measured value to zero, to perform pseudo diffe-
rential measurements with respect to a reference value; with two sensors and
the same measured variables genuine differential measuring operations can
be performed.

PROCESS FLOW PROGRAMMING

To digitally record the measured values of all connected sensors a cyclic mea-
suring point scan using data acquisition software or an internal time-based
process flow control is required.  For this purpose a software clock and the
print cycle are available.  The measuring operation can be started and stopped
via the interface.

Time and date

When the device is switched off, the date and time-of-day are lost.  For the
purposes of logging a measuring operation these must be reset. 

Print cycle

The print cycle can be programmed to any value between 00:00:01 seconds
and 59:59:59 hh/mm/ss.  The print cycle permits cyclic output of measured va-
lues to the interface.

Print cycle factor

To avoid excessive data flow the data output of certain channels can, as and
when necessary, by means of the print cycle factor, be restricted or altogether
disabled.  It is also possible to summate pulses over longer cycles. 

Output

All measured values and programming values are accessible via various inter-
face cables with RS232, RS422, or optic fiber interface.  The output of measu-
red data can be selected in a list, in columns, or in spreadsheet format.  Files
in spreadsheet format can be processed by any standard spreadsheet soft-
ware.  The print header can be programmed specifically to the company or
application.

Networking

The normal networking possibilities of ALMEMO

®

 devices are restricted in this

case because the device address 0 cannot be modified.  

Software

Each ALMEMO

®

 Manual is accompanied by the AMR-Control software packa-

ge, which can be used to configure the measuring instrument and to program
the sensors.  Measuring operations can also be performed online using the in-
tegrated terminal.  The WINDOWS

®

 software package, WIN-Control, can be

used for data acquisition with graphical presentation and complex data proces-
sing.

    ALMEMO

 2390-3

7   

  Introduction   

Summary of Contents for ALMEMO 2390-3

Page 1: ...Operating instructions Universal measuring instrument ALMEMO 2390 3 V2 0 11 12 2003 ALMEMO 2390 3 1...

Page 2: ...nal voltage supply 11 3 3 Switching ON and OFF 11 4 CONNECTING THE SENSORS 12 4 1 Sensors 12 4 2 Measuring inputs and additional channels 13 5 DISPLAY 14 6 SENSOR PROGRAMMING 15 6 1 Measuring range 15...

Page 3: ...8 TROUBLE SHOOTING 21 9 ELECTROMAGNETIC COMPATIBILITY 22 APPENDICES 23 Technical data 23 Product overview 23 Your contact partners 24 ALMEMO 2390 3 3 Table of contents...

Page 4: ...Interface modules RS232 optic fiber Centronics Manual Section 5 2 The entire ALMEMO networking system Manual Section 5 3 All functions and their operation via the interface Manual Section 6 Complete...

Page 5: ...ALMEMO measuring instrument and to interchange these without the need for any extra settings Function channels Maximum minimum and average values and differences at certain measuring points can be pr...

Page 6: ...easured value for the selected measuring point is shown continuously with auto zero and as and when necessary with measured value correction With most sensors sensor breakage is detected automatically...

Page 7: ...ctor To avoid excessive data flow the data output of certain channels can as and when necessary by means of the print cycle factor be restricted or altogether disabled It is also possible to summate p...

Page 8: ...minimum value HOLD Freeze measured value CLEAR Set measured value to zero Delete Max Min Hold 7 Battery compartment rear of dev 9 volt alkaline manganese battery 6F22 Space for reserve battery 5 LCD...

Page 9: ...s M0 and M1 and then select measuring point M2 see 7 5 8 Evaluating a measuring operation call up maximum minimum values by pressing MAX or MIN key 6 delete maximum and or minimum values by pressing C...

Page 10: ...n by the arrow as shown in the illustrations 2 Connect the battery using the clip Accidental re verse polarity is prevented by the shape of the connections 3 Use the second battery compartment to stor...

Page 11: ...being used 3 2 External voltage supply On the right side of the device there is a socket 4 for connecting an external power supply In our range of accessories we offer a mains adapter ZB 2290 NA 12 V...

Page 12: ...s Additional chan nels are available primarily for humidity sensors temperature humidity dew point mixture ratio or flow probes and chemical probes incorporating a tem perature sensor A single sensor...

Page 13: ...However sensors combined within one connector and sensors with their own power supply are electrically intercon nected and must therefore be operated in isolation The volta ge applied to the measuring...

Page 14: ...w 7 V symbol llights up BAT below 6 V L o b A t Checksum error during device calibration Non connected sensors deactivated measuring points Non permitted measuring range Sensor correction or scaling C...

Page 15: ...iCroSil NiSil N ZA 9020 FSN 200 0 1300 0 C NiSi Fe CuNi L ZA 9000 FSL 200 0 900 0 C FECO Fe CuNi J ZA 9000 FSJ 200 0 1000 0 C IrCo Cu CuNi U ZA 9000 FSU 200 0 600 0 C CUCO Cu CuNi T ZA 9000 FST 200 0...

Page 16: ...0 100 0 C H dt Partial vapor pressure cap FH A646 0 0 1050 0 mb H UP Enthalpy cap FH A646 0 0 400 0 kJ H En Humid temperature FN A846 30 00 125 00 C P Ht Relative humidity psychr FN A846 0 0 100 0 H...

Page 17: ...nual Sec 5 1 1 7 1 Selecting the meas value and the measuring point After the device is switched on the MEASURE function is set and measuring point M0 for the first sensor is selected automatically i...

Page 18: ...LUE function Select the maximum value MAX HOLD MIN MAX CORR M DIM The MIN VALUE function Select the minimum value MIN HOLD MIN MAX CORR M DIM Delete the max min values CLEAR Return to the MEASURE func...

Page 19: ...or low the sensor will be adju sted and the correction values will be saved in the connector EE PROM see below Deleting the reference value The reference value for a channel can be deleted by pressing...

Page 20: ...FY A640 O2 101 Example pH probe Zero point adjustment in a buffer solution of 7 00 pH Measured value HOLD MIN MAX CORR M DIM Zero point adjustment CLEAR HOLD MIN MAX CORR M DIM Gain adjustment in a bu...

Page 21: ...r Fluctuating measured values or segment test or the system hangs in mid operation Remedy Check the cabling for inadmissible electrical connections disconnect the external power supply and the analog...

Page 22: ...m IEC 61000 4 4 1995 A1 2000 2kV The following advisory notes must be observed when operating the device 1 If the standard sensor is extended 1 5 meters care must be taken to en sure that the measurin...

Page 23: ...ZB 2290 NA 230 V AC to 12 V DC 200 mA electrically isolated Adapter cable el isol ZB 2290 UK 10 30 V DC to 12 V DC 250 mA Current consumption approx 4 5 mA without input and output modules Housing 18...

Page 24: ...Your contact partners 24 ALMEMO 2390 3 Appendix...

Reviews: