The power indi-
cator is on, and
the nebulizer
does not work
If the low battery indicator lights, it
means that the batteries are low or
the power is insufficient to operate
the device
Replace the batteries and reinstall
the device.
(See page 110)
The membrane of the medication
chamber is damaged
Replace the medication cham-
ber with a new one and fill in the
drugs. (See page 113)
The contacts electrodes on the
medication chamber are contami-
nated with medicine or water
Clean the contacts electrodes from
the remnants of drugs or water and
restart the device. (See page 118)
The electrodes on the nebulizer or medi-
cation chamber are contaminated
Clean the electrodes and restart
the device. (See page 118)
The membrane of the medication
chamber is strongly clogged
If it still not work normally after clean
-
ing, replace the medication chamber
with a new one.
The nebulizer
turns off during
the use
The medication chamber is not
installed up to the stop
Install the medication chamber
correctly and restart the device.
(See page 113)
The micro - USB cable is not prop-
erly connected to the nebulizer
Connect it correctly and restart
the device. (See page 109)
Out of drugs
Add the drugs to the medication
chamber. (See page 113)
The drugs do not enter the
membrane for more than 15 seconds
Place the nebulizer at right angle that
the drugs flow to the mem-brane
The nebuliz-
ing process is
unstable
The nebulizer was shaken when
using
Hold the nebulizer securely in
your hand. (See page 115)
The nebulizer
does not turn off
automatically
when out of the
drugs
The medication chamber is broken
Replace the medication cham-
ber with a new one and fill in the
drugs. (See page 113)
Some solutions for inhalation can lead
to the formation of a large amount of
foam in the medication chamber
Clean the foam and restart the
device. (See page 113)
The contacts electrodes on the
medication chamber are contami-
nated with medicine or water
Clean the contacts electrodes and
restart the device. (See page 118)
The electrodes on the nebulizer or
medication chamber are contami-
nated
Clean the electrodes and restart
the device. (See page 118)
Leakage of
drugs from
the medication
chamber
Damage to the medication chamber
or run out of the silicone gasket
Replace the medication cham-
ber with a new one and fill in the
drugs. (See page 113)
If your nebulizer still does not work properly (go to agu-baby.com).
Summary of Contents for Tomchi N8
Page 1: ...NEBULIZER MESH AGU N8 DE BEDIENUNGSANLEITUNG FR MODE D EMPLOI EN USER MANUAL RU agu baby com...
Page 49: ...49 DE...
Page 141: ...141 EN...
Page 143: ...143 RU AGU N8...
Page 144: ...144 CE 93 42 EEC WEEE 020 2011 BF IP22 2 12 80 2 15...
Page 145: ...145 RU...
Page 146: ...146 AGU N8 24 3 10 23 C 0 35...
Page 147: ...147 RU...
Page 148: ...148 ml 6 4 2 ml 6 4 2 AA 1 5Vx2 USB S M 4...
Page 149: ...149 RU...
Page 152: ...152 4 3 2 3 4 2 AGU N8 M1 g 7392 59 122 47 2 297 90 15 18 3 1746 5 31 84 42607 41 122 47...
Page 154: ...154 ml 6 4 2 ml 6 4 2 AA 1 5Vx2 Micro USB S M 4...
Page 155: ...155 RU 161 165 161 165 164 165 I O 170...
Page 156: ...156 15...
Page 157: ...157 RU 1 a 1 1 b 1 2 c 1 3 1 1 1 2 1 3...
Page 158: ...158 2 161 165...
Page 159: ...159 RU 5 1 2 a...
Page 160: ...160 3 2 AA LR6 6 3 10 2 30 I O 1 2 I O 1 a 2...
Page 161: ...161 RU 1 1 1 1 2 a 1 1 PUSH 1 2 PUSH...
Page 162: ...162 2 0 5 8 3...
Page 163: ...163 RU 0 9 170 1 1 1 2 1 a I O 1 1 1 1 2 1 15 1 15 30...
Page 164: ...164 2 C I O 170 15 15...
Page 165: ...165 RU 3 1 15 1 1 100 100 2 5 3 0 5 3...
Page 166: ...166 6 5 3 2 2 1 2 2 2 3 164 2 2 2 3 2 1...
Page 167: ...167 RU 3 4 5...
Page 168: ...168 6 1 2 1 2 7...
Page 169: ...169 RU 100 3 12 30 12...
Page 170: ...170 1 PUSH PUSH 1 1 1 2 a C 1 1 PUSH 1 2...
Page 171: ...171 RU 2 2 2 2 3 2 1 a 2 1 2 2 2 3...
Page 172: ...172 1 2 3...
Page 173: ...173 RU...
Page 174: ...174 153 15 C 146 163 155 163 1 158 157 146 155 163 155 155 Micro USB 155...
Page 175: ...175 RU 155 158 155 163 163 158 Micro USB 155 158 15 146 160...
Page 176: ...176 158 159 163 163 158 agu baby com...
Page 177: ...177 RU 20 70 c 75 10 40 15 93 20 70 93 2...
Page 178: ...178 2012 19 EU WEEE Waste Electrical and Electronic Equipment...
Page 179: ...179 RU CISPR 11 1 CISPR 11 61000 3 2 61000 3 3...
Page 182: ...182 1 80 800 2 a FM AGU N8 AGU N8 AGU N8 150 80 3...
Page 184: ...184 CISPR 11 1 CISPR 11 B 61000 3 2 A 61000 3 3 60601 ESD 61000 4 2 6 8 6 8 30...
Page 187: ...187 RU 1 80 800 2 a FM AGU N8 AGU N8 AGU N8 150 80 3...
Page 189: ...189 RU 24 6 micro USB S M...
Page 190: ...190 micro USB S M...
Page 191: ......
Page 192: ......
Page 196: ...AGU is the registered trademark by Montex Swiss AG Tramstrasse 16 CH 9442 Berneck Switzerland...