PE-05 N95 RESPIRATOR DATASHEET/ USER INSTRUCTION
Date:
01/07/22
Document Number: PP004
Revision:
E
Product name: N95 Particulate Respirator
Model Number: PE-05
Part Number: PE-05 S (small size head strap length); PE-05 L (large size head strap length)
Product description
: N95-Particulate Respirator (filter efficiency level of 95% or greater against
particulate aerosols free of oil). Time use restrictions may apply. This product contains no
components made from natural rubber latex. This respirator is disposable and non-sterile.
Important
:
This respirator helps protect against certain particles.
Misuse may result in sickness or death
. For
correct use, consult your supervisor and these user instructions.
Before use, wearer must read and understand these user instructions. Keep these instructions for
reference.
For more information and assistance including technical support, contact us at:
Phenotype Pharmaceuticals
Email: [email protected]
1255 Park Ave, Suite A
Phone: 510-817-4008
Emeryville, CA 94608
Fitting Instructions:
Follow these instructions each time a respirator is worn
:
Clean your hands with soap and water or hand sanitizer before touching the respirator.
Remove a respirator from the package, unfold it and make sure there are no obvious tears or holes
in either side of the respirator or on head straps.
Orient the respirator with the Nose Strip pointing up. The side of the respirator that has a stiff
bendable edge is the top and is meant to mold to the shape of your nose.
Hold the respirator against your face (covering nose, mouth and chin) with one hand, and using
your other hand, position top head strap over the top of your head, above the ears.
For the bottom head strap, use one hand to press against the attachment points where the strap is
connected to the respirator body, and use the other hand to pull the bottom strap over your head to
your neck area, below your ears.
Make sure the respirator covers your nose and wraps under your chin with the nose piece up.
Using the fingers of both hands, bend the metal nose clip around your nose to form a tight seal.
Using only one hand to mold the nose piece may cause an improper fit.
Perform a fit check. Place both hands over the respirator. Exhale sharply to check for air leaks.
If air leaks at the respirator edges, re-adjust position of the two head straps along the sides of your
head until maximum fit and seal is achieved. If air leaks around nose re-adjust the nosepiece as
described.