20
SBI 100 User Instructions Manual
4.4 RS-232 Data Commands for SBI Protocol
The RS232 can be set so a bidirectional connection can be established between the
indicator and the host. To establish this connection, set parameter P4 to 7, and
configure setting P5 (baud) and P6 (parity) to host device. Commands can then be sent
from the host to the indicator using the following commands (ensure the letters entered
are in CAPS) (<CR> - Enter)
Key Symbols
<LF>
Line feed
<CR>
Carriage Return
<ETX>
End of text character
<SP>
Space
H1H2H3
3 status bytes
<p>
Polarity character including minus sign for negative weigh
and a space character for positive.
W1-W7
Weight data
<dp>
Decimal point
U1U2:
Unit measure, kg, lb or oz
Command
Action
Response
W<CR>
Takes a reading
over capacity
under capacity
zero point error
reading (kg or lb)
<LF>^^^^^^^^U1U2 U3U4U5<CR><LF> H1H2H3<CR><ETX>
<LF>________U1U2<CR><LF> H1H2H3<CR><ETX>
<LF>-------- U1U2<CR><LF> H1H2H3<CR><ETX>
<LF><p>W1W2W3W4W5W6<dp>W7U1U2<CR><LF>H1H2H3<CR><ETX>
S<CR>
prints status bytes
<LF> H1H2H3<CR><ETX>
Z<CR>
zeros the scale
<LF> H1H2H3<CR><ETX>
T<CR>
sets up a tare
<LF> H1H2H3<CR><ETX>
U<CR>
changes the units
<LF> U1U2<CR><LF> H1H2H3<CR><ETX>
L<CR>
activates the hold function <LF> H1H2H3<CR><ETX>
X<CR>
switches off the scale
Indicator switches off
?
unrecognised command
<LF>? <CR><ETX>
Summary of Contents for Brecknell SBI-100
Page 1: ...SBI 100 User Instructions AWT 35 501281 Issue AA ...
Page 4: ...2 SBI 100 User Instructions Manual Chapter 7 Error Codes 29 ...
Page 8: ...6 SBI 100 User Instructions Manual 1 4 Declaration of Compliance ...
Page 27: ...SBI 100 User Instructions Manual 25 7 8 9 Tare Tare ZERO 0 ...
Page 32: ...30 SBI 100 User Instructions Manual ...
Page 33: ......