28
SBI 100 User Instructions Manual
P15
0,1,2
1
Signal not within power on
zero point
0 = current weight
1 = calibration zero
2 = power off zero point
3 = continuously display “0-----”
P16
0 - 8
6
Zero tracking
0 = 0d AZT off
1 = +/- 0.25d
2 = +/- 0.5d
3 = +/- 1d
4= +/- 1.5d
5= +/- 2d
6 = +/- 3d
7= +/- 4d
8 = +/- 5d
P17
0 - 3
2
Data filter
0 = very weak
1 = weak
2 = standard
3 = strong
P18
0 - 9
1
Weight stability
0 = +/- 0.5d
1= +/- 1d
2= +/- 1.5d
3= +/- 2d
4= +/- 3d
5= +/- 4d
6 = +/- 5d
7= +/- 6d
8 = +/- 7d
9 = +/-8d
P19
0 - 9
1
Overload range full scale
0 = 0%
1 = +9d
2 = 101%
3 = 102%
4 = 405%
5 = 110%
6 = 120%
7 = 150%
8 = 200%
9 = no limitation
Config Menu Avail. Settings Default
Definition
Detailed Setting