Chapter 7 Maintenance 

In order to obtain the best measurement results, it is very necessary to maintain the sensor regularly. 

Maintenance mainly includes cleaning, inspecting damage of the sensor. You can also view the sensor’s 

status during maintenance and inspection. 

7.1 Sensor Cleaning 

It  is  recommended  that  the  sensor  should  be  cleaned  at  regular  intervals  (usually  3  months, 

depending on the site environment) to ensure the accuracy of the measurement. 

Use water to clean the outer surface of the sensor. If there is still debris, wipe it with a damp soft 

cloth. Do not place the sensor in a direct sunlight or near radiation. In the entire life of the sensor, if the 

total sun exposure time reaches to one hour, it will cause the fluorescent cap aging and going wrong, and 

consequently leading to the wrong reading. 

7.2 Inspection on the Damage of Sensor 

According to the appearance of sensor to check if there is damage; if any damage is found, please 

contact after-sales service maintenance center in time for replacement to prevent malfunction of sensor 

caused by water from the damaged cap.   

7.3

 

Preservation of Sensor 

A

When  you  are  not  using  it,  please  cover  the  product’s  original  protective  cap  to  avoid  direct 

sunlight  or  exposure.  In  order  to  protect  the  sensor  from  freezing,  the  DO  probe  should  be  stored  in  a 

place where it will not freeze. 

B

Keep the probe clean before storing it for a long time. Keep the equipment in a shipping box or a 

plastic container with electric shock protection. Avoid touching it with hand or other hard objects in case 

of scratching the fluorescent cap. 

C

It is forbidden that the fluorescent cap is exposed to direct sunlight or exposure. 

Tel

+1 513 204 5837

 

Web

http

s

://www.

aginova.

com

Sentinel (RS-485) DO •

 Operation Manual

Summary of Contents for Sentinel RS-485 DO

Page 1: ...Sentinel RS 485 DO Operation Manual...

Page 2: ...side fixed installation 5 3 1 3 Railing fixed installation 6 3 2 Connection of Sensor 7 Chapter 4 Interface and Operation 7 4 1 User Interface 7 4 2 Parameter Setting 7 Chapter 5 Calibration of Senso...

Page 3: ...aterials SUS316L Ordinary Version Titanium Alloy Seawater Version Up and down cover PPS glass fiber Cable PUR Power Supply 9 28VDC Communication Protocol MODBUS RS485 Storage Temperature 15 60 Operati...

Page 4: ...ely used in the DO online monitoring of the regulating tank biochemical tank and effluent of sewage treatment plant DO online monitoring in waterworks surface water water used in various industrial pr...

Page 5: ...ion diagram of Sentinel RS 485 Dissolved Oxygen Sensor Chapter 3 Installation 3 1 Installation of Sensors Note The protective cap should be removed before use before measurement Do not screw down the...

Page 6: ...4 3 1 1 Quick Dismantling pool side fixed installation Figure 3 Quick Dismantling pool side installation sketch map Tel 1 513 204 5837 Web https www aginova com Sentinel RS 485 DO Operation Manual...

Page 7: ...5 3 1 2 Classic pool side fixed installation Figure 4 Classic pool side fixed installation sketch map Tel 1 513 204 5837 Web https www aginova com Sentinel RS 485 DO Operation Manual...

Page 8: ...6 3 1 3 Railing fixed installation Figure 5 Railing fixed installation sketch map Tel 1 513 204 5837 Web https www aginova com Sentinel RS 485 DO Operation Manual...

Page 9: ...onnect Note Modbus Poll software is a general software that can be downloaded online 4 2 Parameter Setting 1 Click Setup on the menu bar select Read Write Definition and then set the parameters The sl...

Page 10: ...ed oxygen sensor has been calibrated at the factory and if you need to calibrate yourself follow the steps below The specific steps of automatic air calibration are as follows Dry the sensor add a sma...

Page 11: ...ensor and place it in the anaerobic water until the measured data is stable Double click the 06 and a dialog box pops out Enter 27 for Address in the dialog box 12 for Value then click Send Wait 20 se...

Page 12: ...n the following table MODBUS RTU Baud Rate 4800 9600 19200 38400 57600 Data Bits 8 bit Parity Check no Stop Bit 1bit Register Name Address Data Type Register Number Read Write Description Dissolved Ox...

Page 13: ...60s multiples of 2 Air Calibration Step 1 27 Signed 1 W 16 Step 2 27 Signed 1 W 19 It should be cancelled if you don t want to calibrate after the execution of Step 1 Cancel 27 Signed 1 W 21 Zero poi...

Page 14: ...n code 03 for reading register content 04 The length of the returned data is 4 bytes 00 00 40 E0 The slope read is 7 00 analyze 40 E0 00 00 using IEEE 754 CA 7B CRC16 check code 2 Modify Slave Address...

Page 15: ...equipment return 01 10 00 06 00 02 A1 C9 Send command parsing 01 device address 01 10 Function code 16 for writing register content 00 06 The starting register address write is 00 06 00 02 Write 2 reg...

Page 16: ...equently leading to the wrong reading 7 2 Inspection on the Damage of Sensor According to the appearance of sensor to check if there is damage if any damage is found please contact after sales service...

Page 17: ...d when the cap is found severely damaged during the inspection Steps to replace the fluorescent cap Unscrew the old cap from the sensor and then screw the new cap on Note When removing and installing...

Reviews: