12
1
、
Read the slope
Register Name Address Data Type
Register Number
Read/Write
Description
Slope
6
F
2
W/R
Range:0.5-1.5
Send the command
:
01 03 00 06 00 02 24 0A
The equipment return
:
01 03 04 00 00 40 E0 CA 7B
Send command parsing
:
01
:
device address 01
03
:
Function code 03 for reading register content
00 06
:
The starting register address read is 0006
00 02
:
Read 2 registers
24 0A
:
CRC16
check code
The device returns the analysis:
01
:
device address 01
03
:
Function code 03 for reading register content
04
:
The length of the returned data is 4 bytes
00 00 40 E0
:
The
slope read is 7.00
(
analyze 40 E0 00 00 using IEEE 754
)
CA 7B
:
CRC16
check code
2
、
Modify Slave Address
Register Name Address Data Type
Register Number
Read/Write Description
Modify Slave
Address
17
Signed
1
W
Range: 1-254
Send the command
:
01 06 00 11 00 0F 99 CB
The equipment return
:
01 06 00 11 00 0F 99 CB
Send command parsing
:
01
:
device address 01
06
:
Function code 06 for writing register content
00 11
:
The register address of write data is 0017
00 0F
:
Write data content of 0015