17 Technical Information
Window Meanings
238/288
Turbo-V 1KG Rack Controller User Manual / 87-901-045-01
Window Meanings
Tab. 7
N.
Read/
Write
Data
Type
Description
Admitted Values
000 R/W
L
Start/Stop
(in remote mode the window
is read only)
Start = 1
Stop = 0
001 R/W
L
Low Speed (in remote mode
the window is read only)
0 = OFF
1 = ON
(default = 0)
008 R/W
L
Serial, Remote, or Front Panel
control mode
Serial = 0
Remote = 1
Front Panel = 2
(default = 1)
100 R/W
L
Soft Start
(can be written only in Stop
condition)
YES = 1
NO = 0
101 R/W
N
Set Point reference quantity
0 = Frequency
1 = Current
2 = Time
3 =
Normal
(default = 0)
102 R/W
N
Set Point threshold (expressed
in Hz, mA, s, _, mbar)
(default = 720)
103 R/W
N
Set Point delay: time between
the pump start and the set
point check starting (s).
0 to 999999
(default = 0)
104 R/W
L
Set Point Output logic
0 = high level active
1 = low level active
(default = 0)
105 R/W
N
Set point hysteresis
(in % of threshold)
0 to 100
(default = 2)