01-999021-00 A0100
Installation Planning Guide
31
Chapter 4.
Installation Supplies
Sections in this chapter:
•
4.1 “Required Installation Supplies and Equipment” next
•
4.2 “Recommended Installation Supplies and Equipment” page 34
Certain supplies not provided by Varian, such as helium and nitrogen supplies, must be
obtained by the customer before the Varian installation engineer can start the installation.
“Required Installation Supplies and Equipment” on page 31
supplies in detail.
4.1 Required Installation Supplies and Equipment
The installation engineer needs the following non-Varian supplies and equipment during
installation:
•
Electrical power surge protector
•
Liquid helium supply
•
Liquid nitrogen supply
•
Helium gas supply
•
Nitrogen gas supply
•
Face mask and thermal gloves
•
Heat gun
•
Non-magnetic l adder
Locating a reliable local source of liquid helium and nitrogen is particularly important. As
soon as possible after ordering a system, make arrangements for an initial delivery and an
on-going supply of liquid helium and nitrogen.
Electrical Power Strip and Surge Protector
A six-outlet, 115-V or 220-V power strip is required to provide power for the host computer
and peripherals, including plotter and CD-ROM drive. If possible, this power strip should
provide surge suppression to protect the delicate electrical components of the host
computer system. If enough 115-V or 220-V wall outlets are close enough to the host
computer, the power strip is not required. Surge protection, however, is still recommended.