48
User’s and Service Guide U3047-90001
U3047AM12
Cal Kit Operational Check
1-Port Calibration and Verification Procedure
1. Connect the ECal or the mechanical cal kit to Port 1 or the port you are testing. Torque to 8 in-lb. For
further information refer to the Help menu, see
.
2. Perform a 1-Port Calibration on Port 1. On the network analyzer, select
Response
>
Cal > Start Cal >
Calibration Wizard
.
a. If using a mechanical cal kit, select
SmartCal (Guided Calibration)
>
Next
.
b. If using an ECal module, connect the ECal to a network analyzer USB port. Select
Use Electronic
Calibration (ECal)
>
Next
.
3. Continue following the Cal Wizard prompts. In the Select Calibration Ports and ECal Module dialog box
click
Clear All
, then select
Port 1
>
Next
>
Measure
.
Figure 38
1-Port Calibration
4. Continue to follow the prompts until the Calibration Completed dialog box appears.
5. At the Calibration Completed prompt, select
Save As User Calset
, type the name
999.1
. Overwrite the
Calset if it already exists and
Save
.
NOTE
If you do not have a key board, select
Save As User Calset
>
Edit Name
and save as
999.x
.
X is the port number you are calibrating. Use the numeric keypad on the network analyzer
front panel to enter “999.1.”
Figure 39
Calibration Complete
Help