7
Performance Tests
162
Agilent U1602A/U1604A User’s and Service Guide
Figure 7-5
Test connection from calibrator to instrument and digital multimeter
4
Adjust the output of the calibrator and observe the reading display on
the multimeter.
5
Wait a few seconds for the measurement to settle. Compare the
measurement results displayed on the multimeter to the corresponding
test limits shown in Table 7- 3.
6
Continue to check voltage measurement accuracy with the remaining
Volts/div setting values in Table 7- 3.
7
When you have finished checking all of the power supply setting values,
disconnect the power supply from the oscilloscope.
8
Repeat the same verification procedures for channel 2.