![Agilent Technologies U1270 Series Service Manual Download Page 20](http://html1.mh-extra.com/html/agilent-technologies/u1270-series/u1270-series_service-manual_2868312020.webp)
1
Calibration Procedures
Recommended Test Equipment
20
U1270 Series Service Guide
Recommended Test Equipment
The test equipment recommended for the performance
verification and adjustment procedures are listed below
(
Table 1- 1
). If the exact instrument is not available,
substitute with another calibration standard of equivalent
accuracy.
A suggested alternative method is to use the Agilent 3458A
8½ Digit Digital Multimeter to measure less accurate but
stable sources. The output value measured from the source
can be entered into the instrument as the target calibration
value.
Table 1-1
Recommended test equipment
Application
Recommended
equipment
Recommended accuracy requirements
DC voltage
Fluke 5520A
<20% of the U1270 Series accuracy
specification
DC current
Fluke 5520A
<20% of the U1270 Series accuracy
specification
Resistance
Fluke 5520A
<20% of the U1270 Series accuracy
specification
AC voltage
Fluke 5520A
<20% of the U1270 Series accuracy
specification
AC current
Fluke 5520A
<20% of the U1270 Series accuracy
specification
Frequency
Fluke 5520A
<20% of the U1270 Series accuracy
specification
Capacitance
Fluke 5520A
<20% of the U1270 Series accuracy
specification
Duty cycle
Fluke 5520A
<20% of the U1270 Series accuracy
specification
Diode
Fluke 5520A
<20% of the U1270 Series accuracy
specification
Summary of Contents for U1270 Series
Page 1: ...Agilent Technologies Agilent U1270 Series Handheld Digital Multimeters Service Guide...
Page 10: ...X U1270 Series Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK...
Page 14: ...XIV U1270 Series Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK...
Page 16: ...XVI U1270 Series Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK...