5
TriScroll 610 Dry Scroll Vacuum Pump
Installation
Safety
Installation
Startup
WARNING
Do not remove or modify any safety or
insulating equipment from the pump.
To do so may create a serious safety
hazard and may void the warranty.
This pump is designed to pump air only;
it is not designed to pump explosive,
flammable, toxic, or corrosive gases.
They can cause bodily injury, explo-
sion, or fire.
Although the pump can pump trace
particulates normally found in the
atmosphere, it is not designed for pro-
cess solids, chemicals, powders, sol-
vents, condensates, or particulates.
They can damage the equipment,
degrade its performance, or shorten its
useful life.
This pump is designed to pump air only;
it is not designed to pump explosive,
flammable, toxic, or corrosive gases.
They can cause bodily injury, explo-
sion, or fire.
Although the pump can pump trace
particulates normally found in the
atmosphere, it is not designed for pro-
cess solids, chemicals, powders, sol-
vents, condensates, or particulates.
They can damage the equipment,
degrade its performance, or shorten its
useful life.
Install in an area that is not exposed to
rain, steam, or excessive humidity. They
can cause electric shock, short circuits,
and severe bodily injury.
WARNING
Before inspecting or servicing the
pump, be sure the electrical supply is
disconnected.
Consult a qualified electrician when-
ever changing wiring on the pump.
Protect against short circuits by install-
ing a circuit breaker of the proper
capacity.
CAUTION
Consult a qualified electrician when-
ever changing wiring on the pump.
The pump operates in a clockwise
direction when viewed from the motor
end. (Note the arrow on the pump
frame.) Improper rotation can cause
permanent damage to the pump.
NOTE
Check that the inlet screen is installed
in each pump model before beginning
operation.
WARNING
Do not insert a finger or foreign object
in the path of the fan; serious personal
injury may result or the pump may be
damaged.
CAUTION
Do not block the fan ducts because the
pump can become overheated. A pump
surface temperature in excess of 55 °C
(131 °F) is potentially damaging. If such
conditions are observed, turn the pump
off and allow it to cool. Disassemble,
inspect for damage, and repair if neces-
sary.
Operate the pump at the ambient tem-
perature of 5 °C to 40 ° C (41 °F to 104
°F), otherwise damage to the pump or
shortened operating life may result.
Close the isolation valve between the
vacuum pump and the vacuum cham-
ber before startup or shutdown; debris
may be sucked back into the vacuum
chamber.