![Agilent Technologies TPS-mini 9699181 User Manual Download Page 42](http://html1.mh-extra.com/html/agilent-technologies/tps-mini-9699181/tps-mini-9699181_user-manual_2868784042.webp)
Technical Information
J1 – IN-OUT
17
TPS-mini User Manual / 87-901-003-01 (E)
289/318
PIN N.
SIGNAL NAME
INPUT/OUTPUT
14
PROGRAMMABLE ANALOG
SIGNAL (+)
OUT
15
x
GROUND
x
PROGRAMMABLE ANALOG SIGNAL (-)
OUT
Signal Description
START/STOP
: input signal to start or stop the pump. Without the
supplied cover connector the turbomolecular pump automatically
starts, if you want to stop the turbo-pump from outside, you have to
connect the START/STOP (+) signal to the +24 Vdc pin and the
START/STOP (-) signal to the GROUND pin (this is the default
configuration of the cover connector).
INTERLOCK
: input signal to control the pump rotation. With the
supplied cover connector the INTERLOCK (+) signal is connected to
the +24 Vdc pin and the INTERLOCK (-) signal to the GROUND pin; in
this condition the pump is stopped.
SOFT START
: this input is used to provide a "soft start" to the pump;
in this condition the ramp-up time could be up to 20 min.
SPEED SETTING
: PWM input signal to set the pump speed. The
PWM signal characteristics must be the following:
frequency: 100 Hz +/-20 %
amplitude: 24 V max
duty cycle range: from 25 % to 75 %
corresponding to a pump speed from 1100 Hz to 1167 Hz (see the
following diagram).