Tab. 1
FLANGE
FIXING DEVICE
N.
FIXING TORQUE
ISO 63
M10 clamps
4
22 Nm
M8 claws
4
11 Nm
The turbopump with ConFlat inlet flange must be fixed to the
vacuum chamber by means of the appropriate Agilent hardware. See
the appendix "Technical Information" for a detailed description.
NOTE
The TPS-mini cannot be fixed by means of its base.
For installation of optional accessories, see "Technical Information".
WARNING!
The System must be installed in a way that allows an easy interruption of the
line voltage.
WARNING!
Power supply cord: The correct cable for electrical wiring is a three wires
(Ph+N+Earth) cable. The wire section has to be at least 0.75 mm
2
(AWG18).