Pump Module User Guide
Maintenance and troubleshooting
30
Pump Module User Guide
5
Carefully lower the cover to clamp the tube in place.
When the pump head top closes on the tubing, the tubing is clamped and
stretched to locate it in the correct position and with the correct tension.
The correct tension is essential for good tube life.
IMPORTANT
A new Marprene tube in the pump head requires re- tensioning
after the first 30 minutes of running. To re- tension the tubing, open the flip-
top cover of the pump head, allow the tube to settle naturally across the
rollers, and then reclamp the tube.
Reporting problems
If you are unable to resolve a problem, contact Automation Solutions Technical
Support. For contact details, see the
section of the title page.
Make sure you have the serial number of the device when contacting
Automation Solutions Technical Support. You can find the serial number on
the back of the device (
1
).
Related information
100V-240V
50/60Hz
1.5 Amps
00305
SN SGS11LHAB2001
1
For information about...
See...
Power and environmental
requirements
“Laboratory setup requirements” on
page 6
Safety guidelines