![Agilent Technologies Pump Module Original User Manual Download Page 15](http://html1.mh-extra.com/html/agilent-technologies/pump-module/pump-module_original-user-manual_2867788015.webp)
13
Pump Module User Guide
Autofilling options
Pump Module User Guide
Figure
Open Wash Reservoir
When properly configured in the automation software, the Pump Module
automatically fills and drains the reservoir. The Open Wash Reservoir can be
set up to run in either of two modes:
•
Overflow mode
. Fresh wash solution enters the reservoir from the two
inlet ports (
1
), overflows into the overflow trough, and is pumped through
the two outlet ports (
2
) to waste.
•
Fill and empty mode
. Fresh wash solution enters the reservoir from the
two inlet ports (
1
), but is drained via the bottom center port (
3
). In this
mode the reservoir is filled and emptied one or more times between each
wash cycle. The reservoir design ensures maximal emptying to reduce the
contamination load in the wash fluid between each cycle.
Related information
Item
Description
Overflow mode
Fill and empty mode
1
Inlet ports
Connected to input
(fill) tubing
Connected to input (fill)
tubing
2
Outlet ports
Connected to output
(empty) tubing
Plugged
3
Bottom center
port
Plugged
Connected to output
(empty) tubing
1
3
2
For information about...
See...
Setup workflow
Maintaining the Pump Module