Instructions for Use

General Information

This  equipment  is  designed  for  use  by  professionals.  The
user  should  read  this  instruction  manual  and  any  other
additional  information  supplied  by  Agilent  before
operating  the  equipment.  Agilent  will  not  be  held
responsible  for  any  events  that  occur  due  to
non-compliance  with  these  instructions,  improper  use  by
untrained  persons,  non-authorized  interference  with  the
equipment, or any action contrary to that provided for by
specific national standards.
The  TriScroll™  300  is  a  dry,  scroll  vacuum  pump.  This
pump is suitable for pumping air or inert gases. The pump
is  not  intended  to  pump  toxic,  corrosive,  explosive,  or
particulate-forming gases.
The  following  paragraphs  contain  all  the  information
necessary  to  guarantee  the  safety  of  the  operator  when
using  the  equipment.  Detailed  information  is  supplied  in
“Technical Information” on page 5.
This manual uses the following standard safety protocol:

Storage

When  transporting  and  storing  the  pump,  the  following
environmental requirements should not be exceeded:

Unpacking and Inspection

The shipping container is a double carton. 
1. After opening the outer box, remove the foam packing.
2. Slit open the inner box.
3. Lift the pump with the plywood base out of the inner

box.

4. Remove  the  four  bolts  securing  the  pump  frame  to

the plywood base. 

5. Locate the NW16 exhaust fitting and set it aside. 
6. Inspect the pump for damage. 

If there is shipping damage, contact the freight carrier 
and your local Agilent sales office immediately. 

7. Save the carton and packing materials.
Total  weight  of  the  package,  including  the  pump,  is
approximately 33.6 kg (74 lbs).

WARNING

The warning messages are for 

attracting the attention of the opera-

tor to a particular procedure or prac-

tice which, if not followed correctly, 

could lead to serious injury.

CAUTION

The caution messages are displayed 

before procedures, which if not fol-

lowed, could cause damage to the 

equipment.

   NOTE

The notes contain important 

information taken from the text.

Temperature:

–20 °C to 60 °C (–4 °F to 140 °F)

Relative humidity:

0 to 95% (non-condensing)

WARNING

When unpacking the pump, be sure 

not to drop it, and avoid any kind of 

sudden impact or shock vibration 

to it.

WARNING

The TriScroll 300 weighs 26 kg 

(57 lbs). To avoid injury, use proper 

lifting techniques when moving the 

pump. 

   NOTE

Normal exposure to the environ-

ment cannot damage the pump. 

Nevertheless, it is advisable to keep 

the pump inlet closed until the 

pump is installed in the system.

Summary of Contents for PSS06001INV

Page 1: ...INSTALLATION AND OPERATION MANUAL Vacuum Products Division TriScrollTM 300 Inverter Dry Scroll Vacuum Pump Manual No 699904355 Revision C October 2009...

Page 2: ...Copyright 2009 Agilent Inc TriScroll 300 Inverter Dry Scroll Vacuum Pump TriScroll is a trademark of Agilent Inc...

Page 3: ...ce 4 Technical Information 5 Pump Electrical Controller 8 Technical Specifications 9 Serial Interface Specification RS485 and RS232 10 Gas Ballast 12 Bearing Purge 12 Purge Kit 12 Isolation Valve Kit...

Page 4: ...TriScroll 300 Inverter Vacuum Pump This page intentionally left blank...

Page 5: ...known Pump 12 8 Isolation Valve Position 13 9 Pumping Speed Percentage Drop 14 10 VPI and CT100 Kit 15 11 Serial Number Location 15 List of Tables Table Title Page 1 Power Cord Selection 3 2 Circuit B...

Page 6: ...TriScroll 300 Inverter Vacuum Pump This page intentionally left blank...

Page 7: ...ing verwijst aan de volende norm en of richtlijn en beantwoodt a cui se rifersce questa dichiarazione conforme alla e sequente I norma o documento I normativo i We Wir Nous Nosotros Wij Noi Agilent In...

Page 8: ...g representatives of Seller which are inconsistent or in conflict with the terms of this warranty shall not be binding upon Seller unless reduced to writing and approved by an officer of Seller Warran...

Page 9: ...rton 1 After opening the outer box remove the foam packing 2 Slit open the inner box 3 Lift the pump with the plywood base out of the inner box 4 Remove the four bolts securing the pump frame to the p...

Page 10: ...qualified electrician whenever wiring the pump CAUTION Although the pump can pump trace particulates normally found in the atmosphere it is not designed for process solids chemicals powders solvents c...

Page 11: ...verheating Circuit Breakers Table 2 lists the circuit breaker capacity required for the pump Locate the switch or circuit breaker near the pump Inverter with P N S4897001 only NOTE The pump must be co...

Page 12: ...tional and local standards NOTE Pump operation is interlock depen dent The provided mating connec tor must be plugged into P1 to start the pump NOTE When the controller is switched on an electronic se...

Page 13: ...ovided Ambient operating temperature 5 C to 40 C 41 F to 104 F Storage temperature 20 C to 60 C 4 F to 140 F Motor rating 0 67 HP 0 5 kW Operating voltage 50 60 Hz 200 240 VAC Single phase 100 120 VAC...

Page 14: ...DRAFT 6 5 11 TriScroll 300 Inverter Vacuum Pump 6 Figure 1 Interface Drawing with Dimensions Figure 2 Pumping Speed Curves...

Page 15: ...g 3 Inlet NW25 4 Inlet Screen 5 NW16 Exhaust Adapter Not shown 6 Bearing Purge Port 1 4 National Pipe Thread 7 Pump Frame 8 Frame Screws M6 4 9 Gas Ballast Port 1 4 National Pipe Thread 10 Mounting Ho...

Page 16: ...l system variables are also viewable Pump operation is interlocked through the P1 connector Pins 4 and 5 and Pins 8 and 9 must be connected together to enable operation A prewired connector is supplie...

Page 17: ...frequency will be the full speed factory set 62 Hz Vin 1 V Driving frequency will be the minimum speed 35 Hz Figure 5 Driving Frequency vs Analog Input Table 4 P1 Interlock PIN SIGNAL NAME IN OUT 1 VP...

Page 18: ...ogical Start Stop 1 START 0 STOP 001 Logical Low Speed 0 OFF 1 ON 008 Logical Remote Serial Configuration 1 REMOTE 0 SERIAL 102 Numerical Set Point Value Hz 105 Numerical Set Point Hysteresis 0 100 10...

Page 19: ...al Environment Temperature C 300 Numerical Cycle Time min 301 Numerical Cycle Number 302 Numerical Pump Life h 319 Alphanumeric Controller Model 323 Alphanumeric Controller Serial Number 325 Alphanume...

Page 20: ...t port is not recommended Bearing Purge A 1 4 National Pipe Thread bearing purge port item 6 on Figure 3 on page 7 protects the main crankshaft bearings in applications where large amounts of water ar...

Page 21: ...or overload protection This VPI kit includes a 24 VDC VPI valve a communication cable a 24 VDC power supply and a splitter cord The order numbers for NW25 and NW40 inlets are VPI25INV24DC and VPI40INV...

Page 22: ...SP1 the inverter drives the pump at full speed Below the lower set point value SP2 the scroll is driven at minimum speed Higher than set point 2 but lower than set point 1 the driving frequency is fu...

Page 23: ...consists of the CT 100 Active Rough Vacuum Gauge Communication Cable 24 VDC Power Supply Power Splitter Cord Tee Centering Ring s Clamp s and a VPI valve Figure 10 Figure 10 VPI and CT100 Kit For the...

Page 24: ...available for Agilent scroll pumps These are used in place of the standard feet installed on the pump for applications that require minimal vibration transmission from the pump to the installation The...

Page 25: ...r cut Repair or replace Excessive voltage drop Check size and length of power supply cable Defective motor Inspect Contact Agilent Poor ultimate pressure System leak Locate and repair leak Water in pu...

Page 26: ...cation requires the pumping of high quantities of water vapor Use of the bearing purge kit PTSPURGEKIT mentioned earlier will keep this water from impacting bearing life Maintenance should be performe...

Page 27: ...d to rebuild TriScroll 300 series pumps All TriScroll 300 Series models PTSS0300MK Maintenance Tool Kit All fixtures and tools required to perform any maintenance on TriScroll 300 Series pumps All Tri...

Page 28: ...ice at 1 800 882 7426 See the back cover of this manual for a listing of our sales and service offices Visit our web site at http www chem agilent com en US Products Instruments vacuum pages default a...

Page 29: ...the core product e g inlet screens vent valves x Prior to shipment drain any oils or other liquids purge or flush all gasses and wipe off any excess residue x If ordering an Advance Exchange product p...

Page 30: ...d or been exposed to any toxic or hazardous materials OR HAS pumped or been exposed to the following toxic or hazardous materials If this box is checked the following information must also be filled o...

Page 31: ...Other Vacuum leak Other Section 6 ADDITIONAL TERMS Please read the terms and conditions below as they apply to all returns and are in addition to the Agilent Technologies Vacuum Product Division Produ...

Page 32: ...TriScroll 300 Inverter Vacuum Pump This page intentionally left blank...

Page 33: ......

Page 34: ...y Agilent Technologies Vacuum Products Division via F lli Varian 54 10040 Leini Torino ITALY Tel 39 011 997 9111 Toll Free 00 800 234 234 00 Fax 39 011 997 9350 Japan Agilent Technologies Vacuum Produ...

Reviews: