Stopping a seal cycle in progress
About this topic
This topic explains how to stop the PlateLoc when a seal cycle is in
progress. You can use this method to stop the device in an emergency.
Standalone device
procedure
To stop a seal cycle that is in progress:
1.
On the
Main Menu
, press
STOP.
To stop a seal cycle in an emergency:
1.
In case the STOP button on the touch screen does not work or is not
responding fast enough, turn off the PlateLoc. The on/off switch is
on the back of the device. Alternatively, you can disconnect the
power to the PlateLoc.
The following diagram shows the location of the on/off switch and
the power connection on the back of the PlateLoc.
Lab automation
system procedure
You can stop a seal cycle from the Velocity11 lab automation system
software. For detailed instructions, see the Velocity11 lab automation
system software user guide.
00206
PlateLoc
screen menu 1
Main Menu
Seal Time: 1.8s
S e t T i m e : 1 . 8 S e c o n d s
S e t Te m p : 1 6 0 D e g r e e s C
A i r p r e s s u r e : 8 9 p s i
Seal Temp: 160°C
Status: Running
Setup
S
T
O
P
Press to stop the seal
cycle in progress.
00191
PlateLoc
connections
00191
PlateLoc
connections
Press to turn off.
Disconnect power.
Summary of Contents for PlateLoc
Page 1: ...Agilent Technologies PlateLoc Thermal Microplate Sealer User Guide...
Page 4: ......
Page 54: ...Chapter 2 Unpacking and installation PlateLoc User Guide 40...
Page 78: ...Chapter 3 Getting started PlateLoc User Guide 64...
Page 90: ...Chapter 4 Sealing plates PlateLoc User Guide 76...
Page 130: ...Appendix A Integrating the PlateLoc into third party systems PlateLoc User Guide 116...
Page 138: ...Appendix C Glossary PlateLoc User Guide 124...
Page 142: ......
Page 143: ......