Maintenance
Testing Solvent Level Sensors
Testing Solvent Level Sensors
If your PL-GPC 220 solvent module is fitted with steel solvent tanks a pair of
liquid level sensing float switches will have been provided. These can be tested
after removing them from the tanks and laying them on the floor of the solvent
compartment.