![Agilent Technologies PL-GPC 220 User Manual Download Page 64](http://html1.mh-extra.com/html/agilent-technologies/pl-gpc-220/pl-gpc-220_user-manual_2867785064.webp)
64
PL-GPC 220 User Manual
6
Error Information
Errors
Errors
If an error occurs on the system the error handling system will generate a
dialog similar to that shown below with a description of the error.
Errors are categorized as two types:
•
Warnings
Warning messages simply display the appropriate dialog and the Error
indicator on the front panel is flashed orange.
•
Errors
Errors display the appropriate dialog the error indicator is flashed red and
the beeper is sounded.
Depending on the severity of the error, following a short time delay, the
instrument steps in to render the system safe. Details of the errors reported
and the remedial actions are tabulated below, however, in general, severe
errors will put the instrument into the STANDBY mode and the remainder will
allow the system to continue.
Figure 15
Error Handling Dialog
When an error occurs, pressing the help button on the error dialog will open
the help file at the troubleshooting section giving further instructions on how
to overcome the error reported.