Agilent Technologies PHD-4 Instructions Download Page 3

 

 

 

 

 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

 

 
 
 
 

4.  REFILLING THE HELIUM RESERVOIR 

When the reservoir pressure goes down to 
approx 30psi (2 bar), the capillary leak needs to 
be refilled with pure Helium. 
Refilling steps: 
1)  Attach a clean line from pure helium bottle to 

the fill valve of the leak, but do not tighten 
the connection; 

2) Flow helium through the line to purge the 

remaining air from the line; 

3)  Tighten the connection to the valve; 
4)  Open the fill valve; 
5)  Fill the leak to 11 bar (160 psig); 
6) Loosen the connection from the helium 

source; 

7)  Open the fill valve and vent the leak to 2 bar 

(approximately 30 psig); 

8)  Repeat steps 3-7 three times to assure gas 

purity; 

9)  Fill to the pressure indicated on the label to 

obtain the desired flow rate. 

 

To avoid clogging of the capillary, the line from 
your helium source must be clean and free of 
moisture. Copper or stainless steel lines are 
recommended. A rubber hose should NOT be 
used. 

5. STORAGE 

Capillary leaks are subjected to clogging and 
must be kept clean and dry. Store in an air-tight 
container when not in use. Care should be 
takenthat the end of the capillary tube is not 
exposed to oil and dirt. 
Recalibration of the gauge is recommended on 
an annual or biannual basis depending on the 
severity of the environment in which the leak is 
being used.  

DISPOSAL 

Meaning of the "WEEE" logo found in labels 

The following symbol is applied in accordance with 
the EC WEEE (Waste Electrical and Electronic 
Equipment) Directive. 
This symbol (

valid only in countries of the 

European Community

) indicates that the product it 

applies to must NOT be disposed of together with 
ordinary domestic or industrial waste but must be 
sent to a differentiated waste collection system. 
The end user is therefore invited to contact the 
supplier of the device, whether the Parent Company 
or a retailer, to initiate the collection and disposal 
process after checking the contractual terms and 
conditions of sale. 
 

 

 

 

CAUTION