Agilent Technologies PHD-4 Instructions Download Page 1

87-900-073-01(B) 

April 18

th

, 2011 

 

PHD-4 capillary leak 

Model 969-3540 

 

 

1.

 

GENERAL 

The Agilent PHD-4 capillary leak model no. 969-3540 
has been designed to allow a fast, simple and effective 
way to check the calibration of the Agilent HPD-4 leak 
detector. 
The PHD-4 capillary leak contains a crimped stainless 
steel capillary to regulate the helium flow. The leak 
element is attached to a 110 cc stainless steel reservoir 
with an adapter that mates with the PHD-4 probe. 
The reservoir has a pressure gauge for adjusting the 
driving pressure behind the capillary leak element in 
order to change the helium flow rate. The gauge 
measures the pressure up to 11 bar (160 psig). 
The calibration is obtained by comparing the Helium 
concentration generated by the capillary leak in the 
provided calibration chamber with the actual reading of 
PHD-4 and correcting this one if necessary.  
 
 
 
 

The value of the Helium concentration is of the 
order of 100 ppm (for a reservoir pressure of 100 
PSI, 7 bar), and is specified in the calibration curve 
supplied with each unit. 
 

NOTE 

The supplied calibration curves are valid for PHD-4 
devices  equipped with the standard length probe. 

 

2. DESCRIPTION 

The PHD-4 capillary leak package is shown in 
figure 1; 
it includes the following parts: 
1)  PHD-4 capillary leak; 
2)  Nylon bag containing O-ring Parker 2-120 BUNA 
    (2 pcs.); 
3) Instruction manual. 

 
 
 
 
 
 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

Figure1

 

 
 

INSTRUCTIONS 

Agilent Technologies Inc. 
121 Hartwell Avenue, Lexington, MA 02173, USA 

Agilent Technologies Italia S.p.A. 
Via F.lli Varian, 54 - 10040 Leinì (TO) - Italy 