Getting Started
10
PRBS Generator 10 Gb/s User Guide
In an ESD-safe environment, carefully remove the N4970A from the packaging.
Install on a flat surface with unobstructed air flow to the back panel. Make all
RF connections between the N4970A and the DUT or test equipment BEFORE
applying AC power by plugging the converter module into the N4970A. Also,
remove AC power from the N4970A instrument before disconnecting any RF
connections. Plug the AC power cord into the power converter module and a
wall socket, then plug the converter module into the N4970A.
1.2
Important Notes
•
Use ESD protection at all times when using the system.
•
Review min/max specifications before applying input signals.
•
Use only SMA-connectors on the RF ports.
•
Use dust jackets on unused back panel connectors.
•
Situate the instrument away from heat sources.
1.3
Performance Recommendations
1.
When using differential-mode connections, ensure the cables are
phase balanced.
2.
Differential connectors may be used single-ended if second end
terminated in 50
Ω
.
3.
Use high quality cables and connector savers (or adaptors).
4.
Keep cable lengths short and minimize number of cable bends.
5.
Use a 7 to 10 in-lbs torque wrench when attaching connectors.