Using the Multiport Test System
How to De-Embed Measurement Data
Chapter 3
47
How to De-Embed Measurement Data
To use test fixtures or adapters with the multiport test system and establish different
calibration and measurement reference planes, the system allows the effect of the fixture or
adapter to be removed by de-embedding.
You select de-embedding options during setup of a measurement. The characterization file
contains the S-parameters in Citifile format, but with additional header information.
You must first characterize any adapters, fixtures, or probes to use the de-embedding
function. See
“How to Characterize Adapters” on page 44
for more information.
How to Make Measurements
The multiport test system measurements require the completion of four prerequisite steps:
• Setup of the software and hardware (see
“How to Set Up the Multiport Test System
and
“How to Install and Start the Balanced Measurement
.)
• Determination of your test requirements
• Characterization of any needed adapters (see
“How to Characterize Adapters” on page 44
• System calibration (see
On completion of these steps, simply connect the device under test (DUT) to the system and
click the
Measure
button on the main screen.
Attenuation
Set the attenuator to allow sufficient measurement dynamic range while considering DUT
compression and avoiding over-driving the receiver in the network analyzer. Too much
attenuation limits calibration accuracy. Calibration accuracy can be improved at the expense
of measurement time with a narrower calibration IF bandwidth. This selection is only
available on the N4413A (ATN-4111A) and N4414A (ATN-4111B) systems.
Summary of Contents for N4413A
Page 4: ...iv ...
Page 11: ...1 1 About the Multiport Test System ...
Page 18: ...About the Multiport Test System Available Options and Accessories Chapter 1 8 ...
Page 19: ...9 2 User Interface ...
Page 27: ...17 3 Using the Multiport Test System ...
Page 89: ...79 4 Performing Time Domain Measurements ...
Page 109: ...99 5 Menu Reference ...
Page 163: ...153 6 Maintenance and Troubleshooting ...
Page 172: ...Maintenance and Troubleshooting Contacting Agilent Chapter 6 162 ...
Page 173: ...163 7 Instrument Information ...
Page 208: ...Instrument Information Specifications and Characteristics Chapter 7 198 ...
Page 209: ...199 8 Safety and Regulatory Information ...
Page 218: ...Safety and Regulatory Information Regulatory Information Chapter 8 208 ...
Page 219: ...209 A Other Technical Procedures ...
Page 224: ...Other Technical Procedures IF Gain Adjustment Appendix A 214 ...