Appendix A
67
Caring for Connectors
Principles of Microwave Connector Care
Principles of Microwave Connector Care
Table A-4
Principles of Microwave Connector Care
Handling and Storage
DO
DO NOT
• Keep connectors clean.
• Extend sleeve or connector nut.
• Use plastic end caps during
storage.
• Touch mating plane surfaces.
• Set connectors contact-end down.
Visual Inspection
DO
DO NOT
• Inspect each connector carefully
before every connection.
• Look for metal particles, scratches
and dents.
• Use a damaged connector -
EVER.
Cleaning
DO
DO NOT
• Try compressed air first.
• Clean connector threads.
• Use any abrasives.
• Get liquid onto plastic support
beads.