18
■
RS232 INTERFA C E
The connection details of the instrument, and connecting cables are as follows :
Some (out of spec) serial ports need a 10 K resistor in the D connector across wires 1 & 2.
Wire 1 (screen)
Wire 2
Wire 3
Instrument Jack Plug
Body (Gnd)
Ring (Tx)
Tip (Rx)
9 Pin D connector for RS232
Pin 5 (Gnd)
Pin 2 (Rx)
Pin 3 (Tx)
The RS232 interface can be used to access and control the instrument from an external computer. All functions can be accessed
remotely. See next pages for available software.
The baud rate can be set to: 0.3, 0.6, 1.2, 2.4, 4.8, 9.6, 19.2, 38.4, 56 Kbaud. The default and printer baud rate is 9.6 Kbaud. To set the
baud rate, on the hidden keypad, select -SHIFT- BAUD RATE, then -/+ to SET the rate, followed by set. At turn on, the default is
restored.
Local / remote modes
The hidden keypad enables the user to lock out the front panel controls, by selecting LOCAL or REMOTE.