![Agilent Technologies N3970A Operating And Maintenance Manual Download Page 19](http://html1.mh-extra.com/html/agilent-technologies/n3970a/n3970a_operating-and-maintenance-manual_2868166019.webp)
16
■
S
TORING
& R
ECALLING
R
EADINGS
There are various ways of storing and recording data, giving the
user flexible ways of working. They are summarised here for
your convenience.
Storage of reference values: A separate reference is stored for
each
λ
and can be recalled when the instrument is next turned
on.
Display hold: To hold the display at it’s current value, push
HOLD. This value can then be stored. Push HOLD again to
continue updating the display.
Automatic recording of maximum & minimum limits: To display
the minimum and maximum value over a period, access the
hidden keypad and push MAX/MIN. The display will show the
maximum and minimum values in sequence. Then push
MAX/MIN for 3 seconds to re-start the recording process. This
function re-sets automatically when instrument parameters are
altered.
Storing & recalling readings using non-volatile memory
In Autotest, stored data includes absolute & relative values at all
possible wavelengths.
Note: To recall data, the unit must be in manual ‘power meter’
mode.
In manual mode, stored data includes absolute, relative and
reference values.
You can store, display or download up to 845 readings in non-
volatile memory.
Memory No
Identifier
Space
1 - 199
mem
199
200 - 1990 (steps of 10)
mem
179
1 - 199
ID
199
200 - 1990 (steps of 10)
ID
179
1 - 199
mem / ID
89
Total memory
845
Summary of Contents for N3970A
Page 1: ...Operating and Maintenance Guide N3970A OPTICAL POWER METER...
Page 2: ......
Page 24: ...N3970A UM 1 ACCURACY CONSIDERATIONS 21...
Page 46: ...43 N3970A UM 1 NOTES...
Page 47: ...44 NOTES...