Step 4: Install the Module

Step 4: Install the Module

PXI hardware does not support "hot-swap" (changing modules while power is applied to the chassis)

capabilities. Before installing or removing a module to/from the chassis, power off the chassis to pre-
vent damage to the module.

This module can be used in a chassis with cPCI(J1), PXI-1, or PXIe hybrid slots.

Recommended Practices for Temperature Control

l

Use slot blockers and EMC filler panels in empty module slots to assure proper operating temperatures.

l

At ambient temperatures above 45° C (113° F) set the chassis fan to High.

l

The use of an Agilent M9018A Chassis and slot blockers optimizes module temperature performance.

Module Installation Procedure

The module can be installed in any standard PXI slot marked with a peripheral slot com-
patibility image (a circle containing the slot number).

The module can also be installed in any hybrid PXI slot marked with a peripheral slot com-
patibility image (the letter “H” and a solid circle containing the slot number).

1. Make sure that the line cord is plugged in to establish earth ground and that the chassis power switch is Off .

2. If the chassis has multiple fan speed settings, ensure that the fan switch is set to

AUTO

.

3. Position the chassis to provide ample space between the chassis fan intake and exhaust vents. Blockage by

walls or obstructions affects the air flow needed for cooling. (Refer to the chassis documentation for more
information about cooling).

4. Before inserting the module into the chassis, back the mounting screws out to ensure that there is no inter-

ference between the screws and the mounting rails.

5. Holding the module by the injector/ejector handle, slide it into an available PXI (or hybrid) slot, as shown in

the figure below.

Agilent M9302A PXI Local Oscillator Startup Guide

13

Summary of Contents for M9302A

Page 1: ...cillator 3 GHz to 10 GHz Startup Guide NOTICE This document contains references to Agilent Technologies Agilent s former Test and Measurement business has become Keysight Technologies For more informa...

Page 2: ...software is for use in the performance of a U S Government prime contract or sub contract Software is delivered and licensed as Commercial computer software as defined in DFAR 252 227 7014 June 1995...

Page 3: ...ing cables or jumpers installing or removing switching cards or making internal changes such as installing or removing jumpers Do not touch any object that could provide a current path to the common s...

Page 4: ...rdous or toxic substance ele ments are expected to leak or deteriorate during normal use Forty years is the expected useful life of the product This symbol indicates the instrument is sen sitive to el...

Page 5: ...n Procedure 12 Step 4 Install the Module 13 Recommended Practices for Temperature Control 13 Module Installation Procedure 13 Agilent M9302A PXI Local Oscillator Front Panel Features 15 Front Panel Co...

Page 6: ......

Page 7: ...Documentation Map Documentation Map Agilent M9302A PXI Local Oscillator Startup Guide 7...

Page 8: ...e CD and at www agilent com find M9302A go to Document Library Manuals l M9302A Soft Front Panel help system l M9302A device driver API help systems IVI C IVI COM and LabVIEW G l M9302A specifications...

Page 9: ...cking a module inspect it for any shipping damage Report any damage to the shipping agent imme diately as such damage is not covered by the warranty see warranty information at beginning of this docum...

Page 10: ...es contained on the CDs are available for download at www agilent com find M9302A Step 3 Install the Software System Requirements Topic Windows 7 and Vista Requirements Windows XP Requirements Operati...

Page 11: ...interface cable you must power up the chassis BEFORE you power up the PC When you power down your chassis Shut Down the PC BEFORE you power down the chassis If you are using an embedded controller co...

Page 12: ...r Signal Analyzer Software and Product Information CD M9392 10002 launch the installer b Follow the installer prompts Choose a Complete installation to install all software and doc umentation or a Cus...

Page 13: ...ity image a circle containing the slot number The module can also be installed in any hybrid PXI slot marked with a peripheral slot com patibility image the letter H and a solid circle containing the...

Page 14: ...the screws are not tightened properly 8 Verify that the PXI chassis fans are operable and free of dust and other contaminants that may restrict airflow 9 Install filler panels and slot blockers after...

Page 15: ...T This SMA female connector is used to deliver the local oscillator sig nal 3 GHz to 10 GHz out of the front panel Front Panel LEDs LED Description STATUS This LED indicates the overall health of the...

Page 16: ...rigid SMA SMA M9361A A4 M9361 20002 M9360A LO 2 OUT M9361A LO IN Cable semi rigid SMA SMA M9351A A5 M9351 20002 M9360A LO 1 OUT M9351A LO IN Cable semi rigid SMA SMA M9351A B1 8120 5016 M9351A REF IN...

Page 17: ...ent IO Libraries Suite Agilent Connection Expert It will display the modules that are installed Review the configuration data and then launch the SFP This will provide control of the module for self t...

Page 18: ...are the signal paths to the 10 MHz REF OUT and the two 100 MHz REF OUT connectors The fol lowing process demonstrates that all associated connectors cables and circuitry are operational Required Hardw...

Page 19: ...f these downconverters 1 Run the M9302A soft front panel SFP 2 Conduct a Self Test Utility Self Test Run Self Test a If self test passes go to next step b If self test fails the module needs repair Se...