3
Column Modules
User Guide
25
3
Begin to swage the ferrule to the column using a 1/4-inch open-ended wrench.
The design of UltiMetal Plus Flexible Metal ferrules reduces the probability of column
breakage and minimizes the damage to the thread of the corresponding fittings.
Unlike other ferrule designs, a reliable leak-free seal is best made by tightening the
internal nut to a recommended number of degrees, not by applying more force. Applying
excessive force will not provide a better seal. Also, compressing the flexible ferrule less
minimizes damage to fittings.
To swage the ferrule to the column:
a
Find your ferrule type in
and note the range of degrees to tighten the internal
nut.
b
Tighten the internal nut clockwise to the lower end of your ferrule’s range.
c
Check if the ferrule is gripping the column. If yes, stop. If not, continue to tighten the
internal nut in small 5- to 15-degree increments. Check after each increment to see if
the ferrule is gripping the column. Stop as soon as gripping occurs (see
).
Figure 13.
Tightening the internal nut
Table 4
Degrees to tighten the internal nut for proper swaging
Ferrule part number
Degrees to tighten internal nut
G3188-27501
50 – 100 degrees
G3188-27502
30 – 70 degrees
G3188-27503
20 – 50 degrees
G3188-27504
60 degrees
G3188-27505
40 – 90 degrees
G3188-27506
20 – 50 degrees