![Agilent Technologies LTM II Series User Manual Download Page 24](http://html.mh-extra.com/html/agilent-technologies/ltm-ii-series/ltm-ii-series_user-manual_2867438024.webp)
3
Column Modules
24
User Guide
•
UltiMetal Plus Flexible Metal Ferrule appropriate for the column size (see
•
Column cutter, wafer (5181-8836, 4/pk)
Swage the ferrule onto the column
1
Pass the column end through the internal nut, the ferrule, and the swaging wrench
(
).
Figure 11.
Thread the column through the internal nut, ferrule, and swaging wrench
2
Thread the internal nut into the swaging wrench with the column protruding out of the back
of the wrench (see
)
Figure 12.
Thread the internal nut onto the swaging wrench