3
Column Modules
User Guide
23
To Attach a CFT Nut and Ferrule to a Capillary
Column
This procedure is used to attach a CFT nut and ferrule to a capillary column prior to installation
into a CFT fitting or union. If installing the column ends to the union in a transfer line module,
see
“To Connect the Column to the CFT Unions”
instead.
CAUTION
Wear clean, lint-free gloves to prevent contamination of the parts.
Available ferrules
The available UltiMetal Plus Flexible Metal ferrule packages are listed in
Figure 10.
UltiMetal Plus Flexible Metal ferrules
Each UltiMetal Plus Flexible Metal ferrule part number is uniquely designed to prevent
inventory mix-ups, and to help you find the ferrule you need quickly. Color variations between
ferrules are a normal result of the UltiMetal coating.
Tools Required
•
Two 1/4-inch open-end wrenches
•
Swaging nut (G2855-20555) or swaging tool (G2855-60200)
•
Internal nut (G2855-20530)
Table 3
Available UltiMetal Plus Flexible Metal ferrule packages
Item
Part number
Ferrule description
1
G3188-27501
0.1 – 0.25 mm column id, 10/pk
2
G3188-27502
0.32 mm column id, 10/pk
3
G3188-27503
0.45 – 0.53 mm column id, 10/pk
4
G3188-27504
Plug, 10/pk
5
G3188-27505
0.25 – 0.32 mm UltiMetal column id, 10/pk
6
G3188-27506
0.53 mm UltiMetal column id, 10/pk