Commands by Subsystem
5
Agilent InfiniiVision 7000A Series Oscilloscopes Programmer's Guide
585
:TRIGger:SPI:PATTern:DATA
Command Syntax
:TRIGger:SPI:PATTern:DATA <value>,<mask>
<value> ::= integer or <string>
<mask> ::= integer or <string>
<string> ::= "0xnnnnnn" where n ::= {0,..,9 | A,..,F}
The :TRIGger:SPI:PATTern:DATA command defines the SPI data pattern
resource according to the value and the mask. This pattern, along with the
data width, control the data pattern searched for in the data stream.
Set a <value> bit to "0" to set the corresponding bit in the data pattern to
low. Set a <value> bit to "1" to set the bit to high.
Set a <mask> bit to "0" to ignore that bit in the data stream. Only bits
with a "1" set on the mask are used.
Query Syntax
:TRIGger:SPI:PATTern:DATA?
The :TRIGger:SPI:PATTern:DATA? query returns the current settings of the
specified SPI data pattern resource.
Return Format
<value>, <mask><NL>
See Also
•
"Introduction to :TRIGger Commands"
•
•