:TRIGger Commands
30
Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide
869
:TRIGger[:EDGE]:REJect
Command Syntax
:TRIGger[:EDGE]:REJect <reject>
<reject> ::= {OFF | LFReject | HFReject}
The :TRIGger[:EDGE]:REJect command turns the low- frequency or
high- frequency reject filter on or off. You can turn on one of these filters
at a time.
•
The high frequency reject filter adds a 50 kHz low- pass filter in the
trigger path to remove high frequency components from the trigger
waveform. Use the high frequency reject filter to remove high- frequency
noise, such as AM or FM broadcast stations, from the trigger path.
•
The low frequency reject filter adds a 50 kHz high- pass filter in series
with the trigger waveform to remove any unwanted low frequency
components from a trigger waveform, such as power line frequencies,
that can interfere with proper triggering.
Query Syntax
:TRIGger[:EDGE]:REJect?
The :TRIGger[:EDGE]:REJect? query returns the current status of the reject
filter.
Return Format
<reject><NL>
<reject> ::= {OFF | LFR | HFR}
See Also
•
"Introduction to :TRIGger Commands"
•
•
N O T E
The :TRIGger[:EDGE]:REJect and the :TRIGger[:EDGE]:COUPling selections are coupled.
Changing the setting of the :TRIGger[:EDGE]:COUPling can change the COUPling setting.
Summary of Contents for InfiniiVision 3000 X-Series
Page 1: ...s1 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide...
Page 30: ...30 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide...
Page 54: ...54 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 2 Setting Up...
Page 68: ...68 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 3 Getting Started...
Page 218: ...218 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 6 Root Commands...
Page 242: ...242 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 8 BUS n Commands...
Page 280: ...280 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 11 DEMO Commands...
Page 288: ...288 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 12 DIGital d Commands...
Page 340: ...340 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 15 FUNCtion Commands...
Page 358: ...358 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 16 HARDcopy Commands...
Page 378: ...378 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 18 MARKer Commands...
Page 446: ...446 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 19 MEASure Commands...
Page 564: ...564 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 23 POWer Commands...
Page 842: ...842 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 29 TIMebase Commands...
Page 1054: ...1054 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 35 Error Messages...
Page 1076: ...1076 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 36 Status Reporting...
Page 1232: ...1232 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide Index...