Agilent Technologies G6600 Replacement Instructions Manual Download Page 3

 

 

3

13.

 

Insert the new 1/8” ceramic tube into the fitting with a new Vespel 
ferrule.  Slide the nut over the tube and tighten it until finger tight.  

Lift 

the ceramic tube up 1 to 2 mm so that it will not be crushed by the 
compression fitting.  

Tighten the nut snugly with the 7/16” wrench.  This 

connection must be fairly tight to ensure that the nut does not loosen 
under heat stress, but excessive tightening can crack the ceramic tube. 

 

14.

 

Reinstall the fiber insulating toroid and the white insulating disk.  
Carefully slide the Burner Cover over the large ceramic tube.  It is often 
useful to sight down the upper fitting to ensure that the ceramic tube is 
properly aligned and will not break. 

Failure to install the insulating disk may lead to leaks in the Burner. 

 

15.

 

Reinstall the four screws at the base of the Burner. 

16.

 

Unscrew the Hydrogen Inlet fitting from the Tee, and remove the 1/16" 
tube and ferrule. 

 

 

 

Note 

Summary of Contents for G6600

Page 1: ...r is extremely hot when running Do not touch the burner until it has cooled down 2 Turn off the vacuum pump 3 Remove the hydrogen line by disconnecting the Hydrogen Inlet Fitting at the top of the Burner Use a second wrench to hold the Adapter Tee to avoid excess torque on the ceramic tubes 4 Disconnect the black PFA transfer line from the Tee of the Burner Interface 5 Unscrew the Tee fitting and ...

Page 2: ...er 9 Remove the four horizontal socket head cap screws holding the Burner Case 10 Carefully lift the Burner Case up off of the base and set it aside 11 Remove the white insulating disk and the fiber insulating toroid from the Burner base 12 Remove the 1 8 ceramic tube by loosening the nut in the Burner base Discard the used ceramic tube and ferrule Remove any debris from the inside of the connecti...

Page 3: ...loosen under heat stress but excessive tightening can crack the ceramic tube 14 Reinstall the fiber insulating toroid and the white insulating disk Carefully slide the Burner Cover over the large ceramic tube It is often useful to sight down the upper fitting to ensure that the ceramic tube is properly aligned and will not break Failure to install the insulating disk may lead to leaks in the Burne...

Page 4: ...nto the Burner tightening finger tight 19 Check the flow of air and hydrogen following the Detector manual instructions 20 Reconnect the Heater Cable Thermocouple and all the tubing Pressure test the system following the instructions in the Detector manual The pressure may change slightly due to variation in tube dimensions 21 Turn on the Interface Controller and allow the Burner temperature to st...

Page 5: ......

Page 6: ...Copyright 2007 Agilent Technologies Inc 2850 Centerville Road Wilmington DE 19808 1610 Printed in the USA Feb 2007 ...