6.
Remove the ferrule attached to the top of the 1/8” ceramic tube.
7.
Disconnect the Heater Cable from the heating element leads.
The heating element leads are very fragile. Do not bend them excessively or they
will be damaged.
8.
Disconnect the thermocouple leads from the Interface Controller.
9.
Remove the four
horizontal
socket head cap screws holding the Burner
Case.
10.
Carefully lift the Burner Case up off of the base and set it aside.
11.
Remove the white insulating disk and the fiber insulating toroid from the
Burner base.
12.
Remove the 1/8” ceramic tube by loosening the nut in the Burner base.
Discard the used ceramic tube and ferrule. Remove any debris from the
inside of the connection.
Screw extractors may be useful in removing old ferrules from Burner fittings.
Note
Note