Agilent Technologies G4319A Installation Manual Download Page 3

Installation Guide: TCD-to-FID Series Connection

3

Preparing the GC

1

Turn off the detectors.

2

Set the oven to 25 °C and turn off all other heated zones.

3

Turn off all carrier and detector gases at their sources.

4

Turn off the main power switch.

5

Disconnect the power cord from the power source. Some 
internal parts carry power if the cord is connected.

6

Remove the column(s) and any associated hardware from 
inside the column oven.

7

Lift the hinged GC top cover at its front edge to expose the 
detector area. Remove the cover by raising it to vertical, 
lifting its left hinge pin from its bracket, and then sliding the 
cover to the left to free its right hinge pin.

Perform this procedure only if necessary to add a new detector, 
or to remove and replace an existing detector. Remove the 
panels indicated in the following steps. Note that if an existing 
detector is being completely replaced by a new detector, 
carefully store the old detector assembly and all associated 
parts in a safe place for possible future use.

1

Remove two screws along the lower edge of the left side (flow 
control) panel. Slide the panel slightly to the rear and lift it 
off.

N O T E

Carrier gas flow must be maintained until the column has cooled to avoid 
column damage.

C A U T I O N

Electronic components can be damaged by static electricity: use a 
properly grounded static control wrist strap when removing the 
electronics panel.

Summary of Contents for G4319A

Page 1: ...ector must be removed and replaced To achieve a standard factory configuration the TCD must be located in the front detector position If you are installing a new FID at this time refer to its installa...

Page 2: ...or assembly Swage nuts and ferrules Tools Electrostatic protection such as a grounded wrist strap T 20 Torx screwdriver 7 16 inch wrench Steps 1 Preparing the GC 2 Raising the TCD Block 3 Adjusting Ve...

Page 3: ...and then sliding the cover to the left to free its right hinge pin Perform this procedure only if necessary to add a new detector or to remove and replace an existing detector Remove the panels indic...

Page 4: ...off 3 Remove the back panel by removing two screws at its bottom edge loosening two screws at its top edge and the lifting off the two top screws 4 Finally remove the rear top cover by removing four s...

Page 5: ...ion Guide TCD to FID Series Connection 5 3 Cap the vent tube at the top of the detector Thermal cover Insulation Delta PRT sensor Vent tube Heater and PRT Vent tube support Retaining screw Detector ba...

Page 6: ...installed into the top of the oven remove the retaining screw securing the vent tube support and remove the support 2 Carefully bend the vent tube downwards to be alongside the detector column make up...

Page 7: ...to pack around the bottom and sides of the detector The goal is to fill all voids around the detector body as it is placed into the cavity 6 Remove caps from the detector base and from the vent tube...

Page 8: ...t and Vespel ferrule to install the 1 4 to 1 8 inch adapter into the FID 3 Again following proper swaging procedures use the 1 8 inch brass tubing nut back ferrule and front ferrule to connect the TCD...

Page 9: ...tion Guide TCD to FID Series Connection 9 4 Restore gas supplies and following leak test procedures described in the Maintenance and Troubleshooting manual check for leaks particularly at all new conn...

Page 10: ...hai 200131 P R China Safety Notices CAUTION A CAUTION notice denotes a hazard It calls attention to an operating procedure practice or the like that if not correctly performed or adhered to could resu...

Reviews: