![Agilent Technologies G3388A Operation Manual Download Page 13](http://html.mh-extra.com/html/agilent-technologies/g3388a/g3388a_operation-manual_2868502013.webp)
13
Agilent Technologies
Agilent G3388A Leak Detector
Operation Manual
4
Power Supply
The Agilent Leak Detector is powered by rechargeable nickel
metal hydride batteries. These batteries are not fully charged
when the detector is shipped from the factory and must be
recharged before the detector can be used. It takes
approximately 3 hours to fully recharge the batteries; when
fully charged, they will provide approximately 5 hours of
continuous service.
Do not use the detector for 30 minutes after recharging to allow
its temperature to stabilize.
A blinking battery display or a “Low Battery Error” message
indicates that the batteries are low and must be recharged with
the attached recharger.
Follow this procedure to replace the leak detector’s batteries:
1
Remove the four screws on the back of the leak detector
using an appropriate screwdriver.
2
Set the screws aside.
C A U T I O N
•
Do not use any devices other than the one supplied with the
detector to recharge the detector’s batteries.
•
Do not use the detector while its batteries are being recharged.
•
Do not use the detector after recharging the batteries until its
temperature has stabilized (approximately 30 minutes).
•
Make sure that the detector’s power is OFF when removing or
attaching the recharger to the detector.
•
Do contact Agilent if the battery lifetime appears to be unusually
short; the batteries may need to be replaced.