Agilent G3388A Leak Detector
Operation Manual
3
Function Settings
Set the range, buzzer, and numeric display as shown in this
illustration.
Standard Display
No settings are required.
Press the ZERO / ENTER key to adjust
the numeric value to "0".
a)Setting
the Range
Select either
Hi or Low
Select either
Step, ON or OFF
Select either
ON or OFF
b)Setting
the Buzzer
c)Setting the
numeric display
: High Sensitivity Range
Hi
: A sensitivity of 1/4 of
the High Sensitivity Range
Low
Step
:The buzzer will alarm when
3 or more indicator lights
are lit.
ON
:The buzzer will alarm when
6 or more indicator lights
are lit.
OFF
:The buzzer will not sound.
When it is set to ON,
the indicator mark
( or ) will be displayed.
Also the leak condition
will be displayed numerically.
Tone 1
10
Selecting the
Buzzer Tone
Press the FUNC key
Press the ZERO / ENTER key
At blinking (Selecting) condition
Returns to the
Standard Display
Digital leak indicator
-99 to +99