Step 2. Start the digital recording thermometer
6
Hybridization Oven Calibration Procedure
14
Close the oven door and allow the
digital recording thermometer
to
record temperature data for at least
3
hours.
Make sure the
hybridization oven rotator rack
is set to rotate during
this time.
15
Remove the
digital recording thermometer
from the oven. Let the oven
continue to operate after you remove the
digital recording thermometer
from the oven.
C A U T I O N
Make sure that the
digital recording thermometer
is secure and will not detach from
the rack. If the
digital recording thermometer
comes loose during use, the data
collected can be inaccurate. A loose
digital recording thermometer
that becomes
trapped and prevents the rotator rack from rotating can damage the oven.
W A R N I N G
The
digital recording thermometer
can be hot to the touch. Use caution while
handling.