Appendix: Installing the pneumatics module (early version)
24
5.
Plug the ribbon cable into the front or back connector. Make certain the
connector is firmly seated and locked. If the flow manifold is installed
and plugged into the correct position, the ribbon cable will retain its
fold as shown in the diagram. If you unplugged an inlet ribbon cable in
Step 2, replace it.
6.
From the front of the GC, secure the flow manifold to the carrier by
tightening the captive screw until snug using a T-20 Torx screwdriver.
Folded ribbon cable
Open connector
Locked connector
Front
captive screw
Back
captive screw