Connecting the detector
16
2.
Attach the filament leads to the TCD board in the two left-most
positions of the connector block. Insert a small flat-blade screwdriver
in the rectangular slot and depress it while you insert the filament lead
in the round hole. Polarity does not matter.
Caution
If the
∆
PRT is not properly connected to the detector interface card, filament
burnout can occur.
3.
Similarly, connect the
∆
PRT leads to the two right-most positions in the
connector block. Polarity does not matter.
Filament leads
TCD board
Connector block
∆
PRT leads
(thick, black)
(thin, with pins)
Connect
∆
PRT
Connect Filament