FID Maintenance Kit Instructions
7
Turn over the collector assembly and install the new silicone
rubber gasket.
8
Make sure that the gasket lays flat against the bottom of the
collector.
9
Install the new ignitor with the copper washer installed as
shown below.
10
Tighten the ignitor with a 5/16-inch wrench.