14
Agilent 1200 Series Valves
1
Installation and Configuration
Software Configuration
Software Configuration
The Agilent 1200 Series valves can be controlled by the Agilent ChemStation
Rev. A.09.03 or higher.
C A U T I O N
The maximum number of all modules (including CAN-slave valves) in your Agilent 1200
Series LC system must not exceed 13.
Contact your local sales and service representative to verify, if your configuration is
supported. This is especially important, if the number of module is close to 13.
To configure the Agilent 1200 Series valves
Step
Note
1
Start the Agilent ChemStation
Software.
2
Select
Configure 1200 Access
in the
Instrument
menu.
3
In the upcoming
Configuration
dialog
box select one or multiple valves and
click
Add.
The selected green valve icon now
moves from the left
Available Modules
panel to the right
Configured Modules
panel as displayed in
.
4
Click
OK
to leave the
Configuration
dialog box.
5
After a
Restart
the Agilent
ChemStation Software a new menu
item
Setup Valve
now appears in the
Instrument
menu.
Summary of Contents for G1157A
Page 1: ...Agilent Technologies Agilent 1200 Series Valves User Manual ...
Page 16: ...16 Agilent 1200 Series Valves 1 Installation and Configuration Software Configuration ...
Page 24: ...24 Agilent 1200 Series Valves 2 Operation Handheld Controller functions ...
Page 52: ...52 Agilent 1200 Series Valves 4 Valve Applications Solvent Selection ...
Page 60: ...60 Agilent 1200 Series Valves 5 Parts ...
Page 74: ...74 Agilent 1200 Series Valves 7 Safety Information Agilent Technologies on Internet ...
Page 77: ......