Chapter 3
93
Measurement Synchronization with LXI-B
Configuring LXI Triggers
Configuring an LXI State Trigger
Recalling a state in response to a peer to peer message is done in a
similar way, although only a SCPI interface is available.
Note that the Input Event Location parameter is also a part of the
instrument’s state. This means that each saved state may point to new
states to be recalled on subsequent peer to peer messages. For example,
in
“Configuring an LXI State Trigger”
, Task 1 may include setting the
Location parameter for LAN1 to “c:\States\anotherState.state”. After
4. Monitor Response
This may be done
assuming
an LXI Trigger Using
SCPI” on page 90
or
“Configuring an LXI
Trigger Using the Front
Panel” on page 92
are
complete.
Task
Action
Notes
1. Place the
instrument
in the state
to be recalled
Various SCPI commands may be used.
This step is up to the user’s
discretion
2. Save the
state
:MMEM:STOR:STATe
“c:\States\myState.state”
The path must exist, and is up to
the user’s discretion.
3. Place the
instrument
in a new
state
Various SCPI commands may be used
This step is up to the user’s
discretion
4. Set up the
state trigger
:LXI:EVENt:INPut:LAN:LOCation
“LAN0”,“c:\States\myState.state”
Parameter one may be any Event
Name reported by the query
:LXI:EVENt:INPut:LAN:LIST?
Parameter 2 must match the
path used in Task 2.
5. Enable the
state trigger
:LXI:EVENt:INPut:LAN:ENABled
“LAN0”,1