Chapter 1
61
Introduction to Programming X-Series Applications
STATus Subsystem (Remote Only Commands)
Questionable Integrity Signal Enable
This command determines which bits in the
Questionable Integrity Signal Condition Register will set bits in the Questionable
Integrity Signal Event register, which also sets the Integrity Summary bit (bit 9) in the
Questionable Register. The variable <integer> is the sum of the decimal values of the bits
you want to enable.
Questionable Integrity Signal Event Query
This query returns the decimal value of the sum
of the bits in the Questionable Integrity Signal Event register.
NOTE
The register requires that the associated PTR or NTR filters be set
before a condition register bit can set a bit in the event register. The
data in this register is latched until it is queried. Once queried, the
register is cleared.
Questionable Integrity Signal Negative Transition
This command determines which bits in
the Questionable Integrity Signal Condition register will set the corresponding bit in the
Questionable Integrity Signal Event register when the condition register bit has a
negative transition (1 to 0). The variable <integer> is the sum of the decimal values of the
SCPI Status Bits/OPC
Dependencies
Sequential command
Mode
All
Remote Command
:STATus:QUEStionable:INTegrity:SIGNal:ENABle
<integer>
:STATus:QUEStionable:INTegrity:SIGNal:ENABle?
Example
STAT:QUES:INT:SIGN:ENAB 4 Burst Not Found will be
reported to the Integrity Summary of the Status Questionable
register.
Preset
32767
Min
0
Max
32767
SCPI Status Bits/OPC
Dependencies
Sequential command
Mode
All
Remote Command
:STATus:QUEStionable:INTegrity:SIGNal[:EVENt]?
Example
STAT:QUES:INT:SIGN?
Preset
0
SCPI Status Bits/OPC
Dependencies
Sequential command