Agilent Technologies ESA E4401B Installation Note Download Page 5

Installation Note E4411-90051

 5

Installation Procedure

WARNING

Before you disassemble the instrument, turn the power switch OFF and
unplug the instrument. Failure to unplug the instrument can result in
personal injury.

CAUTION

Electrostatic discharge (ESD) can damage or destroy electronic
components. All work on electronic assemblies should be performed at a
static-safe workstation. Refer to the documentation that pertains to your
instrument for information about static-safe workstations and ordering
static-safe accessories.

To avoid potential RFI leakage and prevent connector damage, tighten screws and RF coax cable
connectors to the following values:

Table 3. Torque Settings

Item

Torque

in-lb

Torque

N-cm

SMA connectors

8.5

95

SMC connectors

5.5

62

3-mm, T-10 TORX screws

9.0

101

3.5-mm, T-10 TORX screws

14

157

4-mm,T-15 TORX screws

21

236

Pozidrive Screws

14

157

Summary of Contents for ESA E4401B

Page 1: ...rinted in USA December 1999 Supersedes P N E4411 90020 Installation Note Agilent Technologies ESA E4401B E4411A E4411B Spectrum Analyzers 1 5 GHz Tracking Generator Retrofit Kit 50 Ω Retrofit Kit Option 1DN 75 Ω Retrofit Kit Option 1DQ ...

Page 2: ...ard to this material including but not limited to the implied warranties of merchantability and fitness for a particular purpose Agilent Technologies shall not be liable for errors contained herein or for incidental or consequential damages in connection with the furnishing performance or use of this material Copyright Agilent Technologies 1999 ...

Page 3: ...01 60176 Products Affected E4401B E4411A E4411B Serial Numbers US00000000 US99999999 To Be Performed By x Agilent Service Center x Personnel Qualified by Agilent x Customer Estimated Installation Time 1 Hour Estimated Verification Time 2 Hours Item Quantity Description Part Number 1 1 Board Assembly Tracking Generator 50 Ω E4401 60128 2 1 Adapter Type N m to STET f 50 Ω 1250 0780 3 4 Screw M3 x 8 ...

Page 4: ...ower meter RF Power sensor Power sensor 75 Ω Option 1DQ only Digital voltmeter Option 1DN only BNC f to dual banana plug Option 1DN only Cable BNC 91 cm 36 in Adapter Type N tee m f f Option 1DN only Adapter Type N m to BNC f Option 1DN only Adapter BNC m to Type N f 75 Ω 2 required Option 1DQ only Cable Type N 62 cm 24 in Cable BNC 23 cm 9 in Cable BNC 122 cm 48 in Torque driver Torque driver req...

Page 5: ...should be performed at a static safe workstation Refer to the documentation that pertains to your instrument for information about static safe workstations and ordering static safe accessories To avoid potential RFI leakage and prevent connector damage tighten screws and RF coax cable connectors to the following values Table 3 Torque Settings Item Torque in lb Torque N cm SMA connectors 8 5 95 SMC...

Page 6: ...ly put the analyzer on the work surface with the front frame 2 and 3 facing down 3 Position the handle 4 as shown in the figure 4 Remove the four screws 5 that hold the rear frame 6 and outer cover 7 in place 5 Lift the instrument rear frame 6 and outer cover 7 up and off of the instrument and set them aside Figure 1 Remove Rear Frame and Outer Cover ...

Page 7: ...tion Note E4411 90051 7 Remove Chassis Cover 6 Refer to Figure 2 Remove the screws 2 and 3 attaching the chassis cover 1 to the chassis 7 Remove the chassis cover from the chassis Figure 2 Remove Chassis Cover ...

Page 8: ...on each side of the instrument by sliding the front frame back toward the instrument a small amount and then gently pulling out on the latch tabs 3 11 Slide the front frame forward just far enough to reach the ribbon connector attached to J2 on the front panel keyboard printed circuit board Disconnect the ribbon cable from J2 12 Remove the front frame and position it with the display down With a f...

Page 9: ...e these screws to 101 N cm 9 in lb Figure 4 Remove the IF Bracket and Install the Tracking Generator Replace Front Frame 16 Make sure the water seal is still in place around the output connector hole on the front panel assembly 17 Carefully slide the front frame onto the chassis and attach the ribbon cable to connector A2 on the front panel keyboard printed circuit board Make sure that the ribbon ...

Page 10: ... clips onto the rear frame can come loose When replacing the rear frame make sure the gasket is in the correct position 26 Replace the rear frame assembly 6 using the four screws 5 to fasten the rear frame to the instrument Torque these screws to 236 N cm 21 in lb Verify Analyzer Performance ESA Models E4401B E4411B Perform the following performance verification tests found in the Agilent ESA Spec...