Installation Note E4411-90051
5
Installation Procedure
WARNING
Before you disassemble the instrument, turn the power switch OFF and
unplug the instrument. Failure to unplug the instrument can result in
personal injury.
CAUTION
Electrostatic discharge (ESD) can damage or destroy electronic
components. All work on electronic assemblies should be performed at a
static-safe workstation. Refer to the documentation that pertains to your
instrument for information about static-safe workstations and ordering
static-safe accessories.
To avoid potential RFI leakage and prevent connector damage, tighten screws and RF coax cable
connectors to the following values:
Table 3. Torque Settings
Item
Torque
in-lb
Torque
N-cm
SMA connectors
8.5
95
SMC connectors
5.5
62
3-mm, T-10 TORX screws
9.0
101
3.5-mm, T-10 TORX screws
14
157
4-mm,T-15 TORX screws
21
236
Pozidrive Screws
14
157